Clinical Cold Welding of the Modular Total Hip Arthroplasty Prosthesis.

نویسندگان

  • Robert K Whittaker
  • Ahmed M Zaghloul
  • Harry S Hothi
  • Imran A Siddiqui
  • Gordon W Blunn
  • John A Skinner
  • Alister J Hart
چکیده

BACKGROUND A head that is "clinically cold welded" to a stem is one of the commonest reasons for unplanned removal of the stem. It is not clear which hip designs are at greatest risk of clinical cold welding. METHODS This was a case-control study of consecutively received hip implant retrievals; we chose the design of hip that had the greatest number of truly cold-welded heads (n = 11). For our controls, we chose retrieved hips of the same design but without cold welding of the head (n = 35). We compared the clinical variables between these 2 groups using nonparametric Mann-Whitney tests to investigate the significance of differences between the cold-welded and non-cold-welded groups. RESULTS The design that most commonly caused cold welding was a combination of a Ti stem and Ti taper: 11 out of 48 (23%) were truly cold welded. Comparison of the clinical data showed that no individual factor could be used to predict this preoperatively with none of the 4 predictors tested showing any significance: (1) time to revision (P = .687), (2) head size (P = .067), (3) patient age at primary (P = .380), and (4) gender (P = .054). CONCLUSION We have shown that clinical cold welding is most prevalent in Ti-Ti combinations of the stem and taper; approximately 25% of cases received at our center were cold welded. Analysis of clinical variables showed that it is not possible to predict which will be cold welded preoperatively. Surgeons should be aware of this potential complication when revising a Ti-Ti stem/head junction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Total Hip Arthroplasty with Modular Stem for Dysplastic Hips in South Asian Population

Introduction: Optimum component positioning and orientation is required to optimize the functional result during total hip arthroplasty for dysplastic hips. Patients and methods: Sixty-two patients (66 hips) including 33 males and 29 females underwent total hip arthroplasty using modular stem prosthesis at an average age of 40.6 years (range 17 to 49 years). Nineteen hips were clas...

متن کامل

Total Hip Arthroplasty Using S-ROM Prosthesis for Dysplastic Hip

PURPOSE The purpose of this study was to evaluate the clinical and radiological results of total hip arthroplasty using a proximal modular femoral stem in patients who had secondary coxarthrosis associated with a dysplastic hip. MATERIALS AND METHODS Forty-two patients (45 hips) with secondary coxarthrosis were evaluated after undergoing primary total hip arthroplasty using an S-ROM proximal ...

متن کامل

Will the Clinical Outcome of Cemented Modular Neck Total Hip Prostheses be Comparable to Conventional Devices?

INTRODUCTION: Modular neck total hip arthroplasty (THA) devices afford the surgeon increased intra-operative flexibility in terms of the degree of head version and offset, as well as allowing more controlled stem positioning. Strains on the outer cortical bone surface have been shown to be strongly dependent on the orientation of a THA. These strains can be a potential cause for mechanical fail...

متن کامل

Cementless Hip Arthroplasty in Southern Iran, Midterm Outcome and Comparison of Two Designs

Background: Cementless hip prosthesis was designed to provide biologic fixation, without the use of cement. The second generation components have shown more reliable bone ingrowths and survival rates. We are reporting a midterm result of two designs of cementless prosthesis in a unique culture with different social habits and expectations.Methods: 52 primary cementless total hip arthroplasty in...

متن کامل

Revision Total Hip Arthroplasty Using a Modular Cementless Distal Fixation Prosthesis: the Zmr(®) Hip System. Clinical and Radiographic Analysis of 30 Cases.

OBJECTIVE To evaluate the clinical and radiographic results from 30 cases of revision of total hip arthroplasty using a modular cementless distal fixation prosthesis: the ZMR(®) Hip System. METHOD Between July 2005 and December 2008, 30 operations were performed, on 14 men and 14 women. Two male patients had bilateral surgery. The mean age was 59.2 years (29-81 years), with a mean follow-up o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of arthroplasty

دوره 32 2  شماره 

صفحات  -

تاریخ انتشار 2017