Incorporating multi-leaf collimator leaf sequencing into iterative IMRT optimization.
نویسندگان
چکیده
Intensity modulated radiation therapy (IMRT) treatment planning typically considers beam optimization and beam delivery as separate tasks. Following optimization, a multi-leaf collimator (MLC) or other beam delivery device is used to generate fluence patterns for patient treatment delivery. Due to limitations and characteristics of the MLC, the deliverable intensity distributions often differ from those produced by the optimizer, leading to differences between the delivered and the optimized doses. Objective function parameters are then adjusted empirically, and the plan is reoptimized to achieve a desired deliverable dose distribution. The resulting plan, though usually acceptable, may not be the best achievable. A method has been developed to incorporate the MLC restrictions into the optimization process. Our in-house IMRT system has been modified to include the calculation of the deliverable intensity into the optimizer. In this process, prior to dose calculation, the MLC leaf sequencer is used to convert intensities to dynamic MLC sequences, from which the deliverable intensities are then determined. All other optimization steps remain the same. To evaluate the effectiveness of deliverable-based optimization, 17 patient cases have been studied. Compared with standard optimization plus conversion to deliverable beams, deliverable-based optimization results show improved isodose coverage and a reduced dose to critical structures. Deliverable-based optimization results are close to the original nondeliverable optimization results, suggesting that IMRT can overcome the MLC limitations by adjusting individual beamlets. The use of deliverable-based optimization may reduce the need for empirical adjustment of objective function parameters and reoptimization of a plan to achieve desired results.
منابع مشابه
Optimal Sequencing of Dynamic Multileaf Collimators
Dynamic multileaf collimator (DMLC)-intensity modulated radiation therapy (IMRT) is used to delivery intensity-modulated beams using an MLC, with the leaves in motion. DMLC-IMRT requires the conversion of a radiation fluence map into a leaf sequence file that controls the movement of the MLC while the beam is on. It is imperative that the fluence map delivered using the leaf sequence file be as...
متن کاملA Monte Carlo based phase space model for quality assurance of intensity modulated radiotherapy incorporating leaf specific characteristics.
Dose calculations for intensity modulated radiation therapy (IMRT) require an accurate description of the radiation field defined by the multileaf collimator. A previously developed Monte Carlo phase space model has been modified to provide accurate dose verification for IMRT treatments on a Novalis linear accelerator. We have incorporated into the model the effects of the multileaf collimator ...
متن کاملLeaf sequencing algorithms for segmented multileaf collimation.
The delivery of intensity-modulated radiation therapy (IMRT) with a multileaf collimator (MLC) requires the conversion of a radiation fluence map into a leaf sequence file that controls the movement of the MLC during radiation delivery. It is imperative that the fluence map delivered using the leaf sequence file is as close as possible to the fluence map generated by the dose optimization algor...
متن کاملDirect leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery.
PURPOSE The authors propose a novel optimization model for volumetric modulated arc therapy (VMAT) planning that directly optimizes deliverable leaf trajectories in the treatment plan optimization problem, and eliminates the need for a separate arc-sequencing step. METHODS In this model, a 360° arc is divided into a given number of arc segments in which the leaves move unidirectionally. This ...
متن کاملDosimetric characteristics of LinaTech DMLC H multi leaf collimator: Monte Carlo simulation and experimental study
This study evaluated the basic dosimetric characteristics of a Dynamic Multi Leaf Collimator (DMLC) using a diode detector and film measurements for Intensity Modulated Radiation Therapy Quality Assurance (IMRT QA). The EGSnrc Monte Carlo (MC) simulation system was used for the determination of MLC characteristics. Radiation transmission and abutting leaf leakage relevant to the LinaTech DMLC H...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 29 6 شماره
صفحات -
تاریخ انتشار 2002