Role of a binary metallic modifier in the determination of cadmium in graphite furnace atomic absorption spectrometry.
نویسندگان
چکیده
In order to discuss the matrix modifier effect of palladium, iron, and a mixture of palladium and iron for the determination of cadmium in graphite-furnace atomic absorption spectrometry (GF-AAS), we measured the absorption profiles of a cadmium line at various compositions of these elements. Variations in the gas temperature were also estimated with the progress of atomization, by using a two-line method under the assumption of a Boltzmann distribution. The atomic absorption of cadmium appeared on the way of heating from the charring temperature to the atomizing temperature while the gas temperature was still low; it was thus considered that cadmium was atomized through direct conductive heating from the wall of the graphite furnace. Therefore, the effectiveness of modifiers for cadmium would be determined through any reactions on the furnace wall at the programmed charring and atomizing temperatures. The addition of iron, palladium, and an iron-palladium mixture all enhanced the absorption signal of cadmium compared to a pure cadmium sample; however, their effects were different from one another. Each addition of iron or palladium to the sample solution led to an enhancement of the cadmium absorbance, indicating that iron or palladium individually became an effective matrix modifier for the determination of cadmium. However, the addition of palladium was ineffective for the matrix modification in the coexistence of large amounts of iron. Although these phenomena are very complicated, and thus cannot be understood completely, any metallurgical reaction between the constituent elements during heating of the furnace wall, such as the formation of solid solutions and intermetallic compounds, would cause the effect of a matrix modifier in GF-AAS.
منابع مشابه
GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY WITH MAlIlRIX MODIFICATION FOR DETERMINATION OF LEAD IN DRINKING WATER
An amoniumvanadate plus sodium molybdate mixed modifier Graphite Furnace Atomic Absorption Spectrometric method has been developed for determination of traces of lead in drinking water. The concentration of lead in the sample is obtained by direct comparison to linear working graph prepared from aqueous standards in the modifier medium, The results indicate the better performance of the pro...
متن کاملAnalysis of Low Lead and Cadmium Concentrations in Soil Extracts by Graphite Furnace Atomic Absorption Spectrophotometry
An analytical procedure was derived to provide an accurate measurement of low lead (Pb) and cadmium (Cd) concentrations in soil extracts using graphite furnace atomic absorption spectrometry (GFAAS). Based on peak area measurements and with background correction, the use of 2% phosphoric acid as chemical modifier enables the GFAAS to give measurements of Pb and Cd concentration as low as 3 fLg ...
متن کاملInternational Union of Pure and Applied Chemistry Division of Chemistry and the Environment Commission on Oils, Fats and Derivatives * the Determination of Cadmium in Oils and Fats by Direct Graphite Furnace Atomic Absorption Spectrometry
A description is given of the development by collaborative study of a standardised method for the determination of cadmium in crude and refined oils and fats by graphite furnace atomic absorption spectrometry. The procedure is rapid and sensitive allowing determination at levels of 1–10 mg/kg
متن کاملAir-assisted Liquid Liquid Microextraction Combined with Graphite Furnace Atomic Absorption Spectrometry for Preconcentration and Determination of Trace Amount of Co(II) and Ni(II) Ions in Water Samples
A simple, rapid and efficient method has been developed for the extraction, preconcentration and determination of cobalt (II) and nickel (II) ions in water samples by air-assisted liquid-liquid microextraction (AALLME) coupled with graphite furnace atomic absorption spectrometry (GFAAS). In the proposed method, much less volume of an organic solvent was used as the extraction solvent in the abs...
متن کاملMeasurement of cadmium residues in muscle, liver and kidney of cattle slaughtered in Isfahan abattoir using grafite furnace atomic absorption spectrometry (GFAAS): a preliminary study
The aim of the present study was to determine the levels of cadmium in the muscle, liver and kidney of cattle from Isfahan and to compare the results with those reported by other countries and with the maximum acceptable levels for human consumption. Samples of muscle, liver and kidney from 60 animals aged 1–10- year-old were collected from the carcasses slaughtered at Isfahan abattoir. Samples...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Analytical sciences : the international journal of the Japan Society for Analytical Chemistry
دوره 26 7 شماره
صفحات -
تاریخ انتشار 2010