Do NiTi instruments show defects before separation? Defects caused by torsional fatigue in hand and rotary nickel-titanium (NiTi) instruments which lead to failure during clinical use.

نویسندگان

  • N V Murali Krishna Chakka
  • P Ratnakar
  • Sanjib Das
  • Anandamy Bagchi
  • Sudhir Sudhir
  • Lavanya Anumula
چکیده

AIMS Visual and microscopic evaluation of defects caused by torsional fatigue in hand and rotary nickel titanium (NiTi) instruments. MATERIALS AND METHODS Ninety-six NiTi greater taper instruments which were routinely used for root canal treatment only in anterior teeth were selected for the study. The files taken include ProTaper for hand use, ProTaper Rotary files and Endowave rotary files. After every use, the files were observed visually and microscopically (Stereomicroscope at 10×) to evaluate the defects caused by torsional fatigue. Scoring was given according to a new classification formulated which gives an indication of the severity of the defect or damage. STATISTICAL ANALYSIS Data was statistically analyzed using KruskallWallis and Mann-Whitney U test. RESULTS Number of files showing defects were more under stereomicroscope than visual examination. But, the difference in the evaluation methods was not statistically significant. The different types of defects observed were bent instrument, straightening/stretching of twist contour and partial reverse twisting. Endowave files showed maximum number of defects followed by ProTaper for hand use and least in ProTaper Rotary. CONCLUSION Visible defects due to torsional fatigue do occur in NiTi instruments after clinical use. Both visual and microscopic examinations were efficient in detecting defects caused due to torsional fatigue. This study emphasizes that all files should be observed for any visible defects before and after every instrumentation cycle to minimize the risk of instrument separation and failure of endodontic therapy.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Current Strategies in Metallurgical Advances of Rotary NiTi Instruments: A Review

NiTi alloy has been used as the raw material for making endodontic files. The interest in NiTi instrument research has not waned with time. In fact, there is a renewed interest in this field, notably through the application of modern research tools and techniques from other fields, making it a truly cross-disciplinary area of research. In 1988, Walia and colleagues introduced nickeltitanium (Ni...

متن کامل

Current Strategies in Metallurgical Advances of Rotary NiTi Instruments: A Review

NiTi alloy has been used as the raw material for making endodontic files. The interest in NiTi instrument research has not waned with time. In fact, there is a renewed interest in this field, notably through the application of modern research tools and techniques from other fields, making it a truly cross-disciplinary area of research. In 1988, Walia and colleagues introduced nickeltitanium (Ni...

متن کامل

Current Strategies in Metallurgical Advances of Rotary NiTi Instruments: A Review

NiTi alloy has been used as the raw material for making endodontic files. The interest in NiTi instrument research has not waned with time. In fact, there is a renewed interest in this field, notably through the application of modern research tools and techniques from other fields, making it a truly cross-disciplinary area of research. In 1988, Walia and colleagues introduced nickeltitanium (Ni...

متن کامل

PathFiles: A New Rotary Nickel-Titanium Instrument for Mechanical Glide Path and Preflaring

NiTi instrument separation is a serious concern in endodontic therapy. The analysis of NiTi files after torsional fracture reveals that the majority of torsional separations occur in the last mm. of the files and in the files with lower taper and/ or diameter. Consequentely, the tip of the smaller files presents the higher risk of torsional fracture and should be protected using low torque moto...

متن کامل

PathFileTM: A New Rotary Nickel-Titanium Instrument for Mechanical Glide Path and Preflaring

NiTi instrument separation is a serious concern in endodontic therapy. The analysis of NiTi files after torsional fracture reveals that the majority of torsional separations occur in the last mm. of the files and in the files with lower taper and/ or diameter. Consequentely, the tip of the smaller files presents the higher risk of torsional fracture and should be protected using low torque moto...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of contemporary dental practice

دوره 13 6  شماره 

صفحات  -

تاریخ انتشار 2012