Dosimetric Validation of Volumetric Modulated Arc Therapy (VMAT) Using AAPM TG-119 Benchmark Plans in an Upgraded CLINAC 2100CD for Flattening Filter Free (FFF) Photon Beams
نویسندگان
چکیده
Background: Recently we have upgraded our Varian Clinac 2100CD with a 6MV FFF beam, this upgrade being the first of its kind in our country. Even though the dosimetric characteristics of FFF beams have been reported both in experimental and Monte Carlo studies, application in planning and delivery is complex. The aim of this study was to validate the commissioning of upgraded FFF beams dosimetrically using AAPM TG-119 bench mark plans for VMAT and to make a comparison with IMRT plans for both flattened filtered and FFF beams. Materials and Methods: AAPM TG-119 proposes a set of test clinical cases for testing the accuracy of IMRT planning and delivery systems. For these clinical cases we generated four treatment plans using IMRT FF, IMRT FFF, VMAT FF and VMAT FFF on a Varian Clinac 2100CD machine equipped with a millennium 120 MLC in Eclipse treatment planning system. Dose prescription and planning objectives were set according to the TG-119 goals and plans were scored based on planning objectives. Plans were compared using dose coverage, the conformity index and the homogeneity index. Point doses were measured at points recommended by TG-119 using a CC13 ion chamber. Planar dosimetry was accomplished using Imatrix and gamma evaluation was conducted using Omnipro IMRT software. Results: Dose distributions of FFF beam based plans were comparable to FF plans for both IMRT and VMAT. Our planning results matched TG-119 planning results. Measured point doses were within ±2% of planned doses and planar dosimetry gamma values were <1 for >95% of data points for all plans. Conclusion: We found a reduction of 40% treatment time for FFF against FF beams for sliding window IMRT. Upgraded FFF beams were in good agreement with TG-119 benchmark plans and goals.
منابع مشابه
Dosimetric influence of Flattening Filter (FF) and Flattening Filter Free (FFF) 6 and 10 MV photon beams on Volumetric Modulated Arc Therapy (VMAT) planning in case of prostate carcinoma
Background: In the treatment of prostate cancer, radiotherapy is the potential to increase second primary cancers such as bladder and rectal cancers. The reasons for this potential are more monitor units (MUs), therefore a larger total body dose because of leakage radiation, a bigger volume of normal tissue is exposed to lower radiation doses. This study was designed to compare the integral dos...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملFlattening Filter-Free Beams in Intensity-Modulated Radiotherapy and Volumetric Modulated Arc Therapy for Sinonasal Cancer
PURPOSE To evaluate the dosimetric impacts of flattening filter-free (FFF) beams in intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) for sinonasal cancer. METHODS For fourteen cases, IMRT and VMAT planning was performed using 6-MV photon beams with both conventional flattened and FFF modes. The four types of plans were compared in terms of target dose homoge...
متن کاملDosimetric superiority of flattening filter free beams for single-fraction stereotactic radiosurgery in single brain metastasis
For single-fraction stereotactic radiosurgery (SRS) using linac in brain metastases, more accurate treatment delivery with higher tumor absorbed doses and lower absorbed doses to normal tissues remains an enormous challenge. The purpose of this study was to investigate the dosimetric superiority in flattening filter free beams (FFF) for volumetric modulated arc therapy (VMAT) in single brain me...
متن کاملDosimetric Comparison between 6MV Flattened Filter and Flattening Filter Free Photon Beams in the Treatment of Glioblastoma with IMRT Technique: A Treatment Planning Study
Introduction: The present study evaluated the dosimetric comparison between 6MV flattened filter (FF) and flattening filter-free (FFF) photon beams in intensity-modulated radiation therapy (IMRT) technique for the treatment of glioblastoma (GBM) patients. Material and Methods: The present study was conducted on 10 patients with GBM previously...
متن کامل