Short implant to support maxillary restorations: bone stress analysis using regular and switching platform.
نویسندگان
چکیده
PURPOSE The aim of this study was to evaluate stress distribution on peri-implant bone simulating the influence of implants with different lengths on regular and switching platforms in the anterior maxilla by means of three-dimensional finite element analysis. MATERIALS AND METHODS Four mathematical models of a central incisor supported by an external hexagon implant (diameter, 5.0 mm) were created, varying the length (15.0 mm for long implants [L] and 7.0 mm for short implants [S]) and the diameter of the abutment platform (5.0 mm for regular models [R] and 4.1 mm for switching models [S]). The models were created using the Mimics 11.11 (Materialise) and SolidWorks 2010 (Inovart) software. Numerical analysis was performed using ANSYS Workbench 10.0 (Swanson Analysis System). Oblique forces (100 N) were applied to the palatine surface of the central incisor. The bone/implant interface was considered perfectly integrated. Maximum (σ(max)) and minimum (σ(min)) principal stress values were obtained. RESULTS For the cortical bone, the highest stress values (σ(max)) were observed in the SR (73.7 MPa) followed by LR (65.1 MPa), SS (63.6 MPa), and LS (54.2 MPa). For the trabecular bone, the highest stress values (σ(max)) were observed in the SS (8.87 MPa) followed by the SR (8.32 MPa), LR (7.49 MPa), and LS (7.08 MPa). CONCLUSIONS The influence of switching platform was more evident for the cortical bone in comparison with the trabecular bone for the short and long implants. The long implants showed lower stress values in comparison to the short implants, mainly when the switching platform was used.
منابع مشابه
A two-short-implant-supported molar restoration in atrophic posterior maxilla: A finite element analysis
PURPOSE The aim of this study was to investigate the stress distribution of 2-short implants (2SIs) installed in a severely atrophic maxillary molar site. MATERIALS AND METHODS Three different diameters of internal connection implants were modeled: narrow platform (NP), regular platform (RP), and wide platform (WP). The maxillary first molars were restored with one implant or two short implan...
متن کاملEffect of Abutment Angulation and Material on Stress and Strain Distributions in Premaxillary Bone: A Three-Dimensional Finite Element Analysis
Background and Aim: Dental implants with angled abutments are often inserted in the anterior maxillary region due to the status of the residual ridge and aesthetic considerations. The purpose of this study was to assess stress and strain distributions in the premaxillary bone around dental implants by means of finite element analysis (FEA). Materials and Methods: Four three-dimensional (3D) fi...
متن کاملBiomechanical Effects of Platform Switching in Two Different Implant Systems: A Three-Dimensional Finite Element Analysis
OBJECTIVES The purpose of this study was to determine the influence of platform switching on stress distribution of two different implant systems using three-dimensional (3D) finite element models. MATERIALS AND METHODS Six 3D finite element models were created to replicate two different implant systems with peri-implant bone tissue, in which six different implant-abutment configurations were...
متن کاملRegular and platform switching: bone stress analysis varying implant type.
PURPOSE This study aimed to evaluate stress distribution on peri-implant bone simulating the influence of platform switching in external and internal hexagon implants using three-dimensional finite element analysis. MATERIALS AND METHODS Four mathematical models of a central incisor supported by an implant were created: External Regular model (ER) with 5.0 mm × 11.5 mm external hexagon implan...
متن کاملThree-dimensional finite element analysis of platform switched implant
PURPOSE The purpose of this study was to analyze the influence of the platform switching concept on an implant system and peri-implant bone using three-dimensional finite element analysis. MATERIALS AND METHODS Two three-dimensional finite element models for wide platform and platform switching were created. In the wide platform model, a wide platform abutment was connected to a wide platform...