In vitro evaluation of the accuracy of two electronic apex locators
نویسندگان
چکیده
This study involves evaluating the accuracy of two electronic apex locators (EALs), Raypex and Neosono Co-pilot. Ten single-root human anterior teeth were used for the study. The crown was sectioned to gain access to the root canal. For each tooth, the reference (or control) length, corresponding to the actual length, was determined, after which all the teeth were measured independently. The results obtained with each EAL were in turn compared with the corresponding control length. The statistical analysis of the results showed that EAL reliability in detecting the apex varies from 80 to 85% for Neosono systems and 85 to 90% for the Raypex systems. Combined with a high observer concordance, these results suggest that electronic root canal measurement can be an objective and acceptably reproducible technique.
منابع مشابه
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...
متن کاملEvaluation of the accuracy of two apex locators in endodontic treatment and retreatments: an ex vivo study
Introduction: Working length determination is important in successful endodontic treatment and retreatment.This study evaluated the accuracy of two electronic apex locators Root ZX and Raypex®6 (EALs) in determining the electronic working length (EWL) of the root canals in endodontic treatment and retreatment. Materials and Methods: Access cavities were prepared on forty extracted, singl...
متن کاملThe effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants
BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...
متن کاملارزیابی دقت دو دستگاه آپکس یاب الکترونیکی (Electronic Apex Locator) از نسل چهارم در تعیین طول کانال (In-vivo)
Background and Aim: Determining correct working length in root canal therapy is of great importance in successful treatment. The purpose of this study was to compare the accuracy of two electronic apex locators of fourth generation, Raypex4 and Apit7, in determining working length in vivo. Materials and Methods: In this test evaluation study performed at Dental School of Tehran University, a to...
متن کاملComparison of Working Length Determination Using Apex Locator and Conventional Radiography in Curved Canals-an In Vitro Study
Introduction: The exact measurement of the root canal length is of great importance in root canal therapy. Although, determination of the root canal length through radiography is the most common method, recently electronic apex locators are being used to determine working length and decrease the number of radiographs. The purpose of this study was to compare the accuracy of an electronic apex l...
متن کاملComparative Evaluation of Accuracy of Three Electronic Apex Locators in Different Simulated Clinical Conditions- An invitro Study
How to Cite this Article: Taneja S, Kumar M, Sharma SS, Gogia H, et al. Comparative Evaluation of Accuracy of Three Electronic Apex Locators in Different Simulated Clinical ConditionsAn invitro Study. Ann Med Health Sci Res. 2017; 7: 190-194 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to...
متن کامل