A biomechanical analysis of plate fixation using unicortical and bicortical screws in transverse metacarpal fracture models subjected to 4-point bending and dynamical bending test

نویسندگان

  • Eirini Liodaki
  • Robert Wendlandt
  • Klaus Waizner
  • Brigitte E. Schopp
  • Peter Mailänder
  • Felix Stang
چکیده

In the published literature there are controversial data to the biomechanical stability of monocortical comparing to the bicortical fixation of metacarpal fractures. The aim of this study was to compare the biomechanical stability of monocortical and bicortical locking osteosynthesis in quasi-static and dynamic 4-point bending tests of composite third metacarpal bone (4th Gen third metacarpal, Sawbones, Malmö, Sweden) fixed with 7-hole locking plate (XXS System, Biotech-Ortho, Wright, Memphis, TN). The tests to determine quasi-static yield and bending strength as well as fatigue strength were conducted in 4 groups of 10 samples after creating standardized mid-shaft transverse osteotomies using a diamont belt grinder (0.3 mm saw blade). The force applied was the dorsal apex loading, similar to the forces applied to metacarpals during normal finger flexion and extension.In the quasi-static testing, no plate breakage was observed in each group. All metacarpals broke at their thinnest part. The average bending strength of the bicortical samples (10.54 ± 0.998 Nm) was significantly higher comparing to the monocortical samples (8.57 ± 0.894 Nm) (P < .001).In the dynamic loading test, all constructs (8 monocortical samples and 7 bicortical) that failed broke at the osteotomy site and the average fatigue strength did not differ in both groups.Consequently, a unicortical plating method may provide adequate strength and stability to metacarpal fractures based on the results of the cyclical loading representative of in vivo loading.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A biomechanical study comparing plate fixation using unicortical and bicortical screws in transverse metacarpal fracture models subjected to cyclic loading.

The use of bicortical screws to fix metacarpal fractures has been suggested to provide no added biomechanical advantage over unicortical screw fixation. However, this was only demonstrated in static loading regimes, which may not be representative of biological conditions. The present study was done to determine whether similar outcomes are obtained when cyclic loading is applied. Transverse mi...

متن کامل

Unicortical versus bicortical locked plate fixation in midshaft clavicle fractures.

Higher rates of poor outcomes in displaced midshaft clavicle fractures treated nonoperatively have recently been reported. Along with expanding indications for operative fixation and increasing application of locked plate constructs, it is unknown whether complications related to bicortical penetration of the clavicle can be avoided using unicortical fixation. The purpose of this study is to co...

متن کامل

Stability of Unicortical versus Bicortical Metacarpal Fracture Internal Fixation Trial (SUBMIT): study protocol for a randomized controlled trial

BACKGROUND Metacarpal fractures are common, accounting for 40 % of all hand injuries. The use of plates for the fixation of these fractures allows early aggressive hand therapy post-operatively, reducing post-operative stiffness. Traditionally, bicortical fixation is the standard practice, where both dorsal and palmar cortices of the metacarpal are drilled through, with screws engaging both cor...

متن کامل

Single column locking plate fixation is inadequate in two column acetabular fractures. A biomechanical analysis

BACKGROUND The objective of this study was to determine whether one can achieve stable fixation of a two column (transverse) acetabular fracture by only fixing a single column with a locking plate and unicortical locking screws. We hypothesized that a locking plate applied to the anterior column of a transverse acetabular fracture would create a construct that is more rigid than a non-locking p...

متن کامل

Biomechanical Comparison of Bicortical, Unicortical, and Unicortical Far-Cortex-Abutting Screw Fixations in Plated Comminuted Midshaft Clavicle Fractures.

PURPOSE The objective of this study was to assess the biomechanical properties of bicortical locking screws, unicortical locking screws, and unicortical far-cortex-abutting locking screw fixation in a cadaver model of comminuted midshaft clavicle fractures stabilized with a locking plate placed on the superior surface of the clavicle. METHODS Nine pairs of adult fresh-frozen cadaver clavicles...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 96  شماره 

صفحات  -

تاریخ انتشار 2017