Comparison of inflationary non-invasive blood pressure (iNIBP) monitoring technology and conventional deflationary non-invasive blood pressure (dNIBP) measurement in detecting hypotension during cesarean section

نویسندگان

  • Akiko Yamashita
  • Shingo Irikoma
چکیده

Background The Nihon Kohden linear inflationary non-invasive blood pressure (iNIBP) monitoring technology is an oscillometric device that measures blood pressure by detecting oscillations during inflation. Systolic blood pressure can be recorded without overinflating the cuff higher than the true systolic pressure. Thus, total time taken for inflation and deflation is shorter than that by the conventional deflation devices. In this study, the ability of iNIBP to detect maternal hypotension during cesarean section faster than deflationary non-invasive blood pressure (dNIBP) monitoring devices under clinical settings was evaluated prospectively. Methods A prospective study of singleton planned cesarean sections at a tertiary center was conducted from August 2015 to April 2016. The combined spinal-epidural anesthesia (CSEA) technique through a single puncture was performed for cesarean section at the center where the study was carried out. An iNIBP cuff was placed on the same arm as the intravenous line, and a dNIBP cuff was placed on the other arm. Due to left uterine displacement by approximately 10° tilt of OR table, hypotension in this study was defined as systolic pressure of 107 mmHg or less, when measured in the left arm, which was about 10 cm lower, and pressure of 92 mmHg or less in the right arm which was about 10 cm higher. This setup was done to evaluate which device detected hypotension faster under clinical settings. A two-tailed Z test was performed to statistically analyze the difference between iNIBP and dNIBP measurement results. Results One hundred singleton planned cesarean deliveries under CSEA were included after 36 weeks of gestation. Out of the 100, 76 women (76%) experienced maternal hypotension. Of these, iNIBP detected hypotension faster than dNIBP in 47 cases (61.8%). Conclusion It was found that iNIBP detected hypotension faster than conventional dNIBP without compromising the reliability of measurement. This may lead to early treatment of maternal hypotension and prevention of adverse events related to the mother and the fetus.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of direct and indirect blood pressure in patients under esophagectomy surgery: a prospective observational study

Background: Arterial pressure is one of the most important physiological variables and often needs to be monitored repeatedly or continuously in perioperative period. Arterial pressure monitoring is one of the standard monitoring in operating room. During general anesthesia, blood pressure can be measured by using a noninvasive arterial pressure method or continuous invasive arterial pressure b...

متن کامل

Monitoring of hypotension after spinal anesthesia in cesarean section with continuous non-invasive arterial pressure measurement

We compared continuous non-invasive arterial pressure device and intermittent oscillometric arterial pressure measurement in maintaining hemodynamic stability during spinal anaesthesia for cesarean section. One hundred American Society of Anesthesiologists (ASA) I-II parturients undergoing elective cesarean section under spinal anesthesia were included. Based on a random number table, the partu...

متن کامل

Continuous Non-invasive Arterial Pressure Device as an Adjunct to Recognize Fluctuating Blood Pressures during Elective Cesarean Section under Subarachnoid Blockade (sab).

BACKGROUND Measuring non-invasive blood pressure (NIBP) in less than one minute intervals (STAT NIBP measurements) is not always feasible. Therefore, large number of undetectable hypotension episodes can only be recognized with continuous beat to beat monitoring of blood pressure, for example, by continuous non-invasive arterial pressure monitor (CNAP). OBJECTIVE The purpose of the current st...

متن کامل

[Non-invasive monitoring based on bioreactance reveals significant hemodynamic instability during elective cesarean delivery under spinal anesthesia].

BACKGROUND AND OBJECTIVES Blood pressure monitoring offers a limited understanding of the hemodynamic consequences of spinal anesthesia for cesarean delivery. The purpose of this study was to assess, with the aid of a non-invasive cardiac output monitor based on bioreactance, the hemodynamic changes during elective cesarean delivery under spinal anesthesia in which intermittent boluses of pheny...

متن کامل

Non-invasive Optical Techniques for determination of blood Glucose levels: A Review Article

This article reviews the development of non-invasive optical techniques for determination of blood glucose concentrations in diabetic patients. Early diagnosis and daily management are essential for ensuring the healthy life of diabetic patients. The determination of blood glucose concentration with common devices involves the chemical analysis of blood samples, which are obtained by pricking t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2018