Clinical Determination of Target Registration Error of an Image- Guided Otologic Surgical System using Patients with Bone-Anchored Hearing Aids

نویسندگان

  • Ramya Balachandran
  • Robert F. Labadie
  • J. Michael Fitzpatrick
چکیده

Image guidance in otologic surgery has been thwarted by the need for a non-invasive fiducial system with target registration error (TRE) at the inner ear below 1.5mm. We previously presented a fiducial frame for this purpose that attaches to the upper dentition via patient-specific bite blocks and demonstrated a TRE of 0.73mm (±0.23mm) on cadaveric skulls. In that study, TRE measurement depended upon placement of bone-implanted, intracranial target fiducials—clearly impossible to repeat clinically. Using cadaveric specimens, we recently presented a validation method based on an auditory implant system (BAHA System®; Cochlear Corp., Denver, CO). That system requires a skullimplanted titanium screw behind the ear upon which a bone-anchored hearing aid (BAHA) is mounted. In our validation, we replace the BAHA with a fiducial marker to permit measurement of TRE. That TRE is then used to estimate TRE at an internal point. While the method can be used to determine accuracy at any point within the head, we focus in this study on the inner ear, in particular the cochlea, and we apply the method to patients (N=5). Physical localizations were performed after varying elapsed times since bite-block fabrication, and TRE at the cochlea was estimated. We found TRE to be 0.97mm at the cochlea within one month and 2.5mm after seven months. Thus, while accuracy deteriorates considerably with delays of seven months or more, if this frame is used within one month of the fabrication of the biteblock, it achieves the goal and in fact exhibits submillimetric accuracy.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Validation of a Fiducial Frame System for Image-guided Otologic Surgery Utilizing Baha® Bone Screws

Image-guided otologic surgeries performed on healthy patients require non-invasive fiducial systems that achieve submillimetric target registration error (TRE). We have described such a system that attaches repeatably to a subject via a dental bite block [1]. We now describe a new method for the clinical validation of such a system that can be conducted in the laboratory with no intraoperative ...

متن کامل

Submillimetric target-registration error using a novel, non-invasive fiducial system for image-guided otologic surgery.

OBJECTIVE Otologic surgery is undertaken to treat ailments of the ear, including persistent infections, hearing loss, vertigo, and cancer. Typically performed on otherwise-healthy patients in outpatient facilities, the application of image-guided surgery (IGS) has been limited because accurate (<1 mm), non-invasive fiducial systems for otologic surgery have not been available. We now present su...

متن کامل

Bite-Block Relocation Error in Image-Guided Otologic Surgery

Otologic surgery is undertaken to treat ailments of the ear including persistent infections, hearing loss, vertigo, and cancer. Typically performed on healthy patients in outpatient facilities, the application of image-guided surgery has been limited because accurate (<1mm), non-invasive fiducial systems for otologic surgery are not available. We have developed such a system, which repeatably a...

متن کامل

A minimally invasive registration method using surface template-assisted marker positioning (STAMP) for image-guided otologic surgery.

OBJECTIVE A new, minimally invasive registration method was developed for image-guided otologic surgery. We utilized laser-sintered template of the patient's bone surface to transfer the virtual markers to the patient's bone intraoperatively and eliminated the necessity for preoperative marker positioning or additional CT scan. STUDY DESIGN Simulation surgeries and clinical application. SUB...

متن کامل

Image-guided otologic surgery

Application of image-guided surgical systems to otologic surgery has been limited by the need for submillimeter accuracy via a fiducial system that is easily useable (i.e. noninvasive and nonobstructive). To overcome these limitations and demonstrate the feasibility of image-guided otologic surgery, a novel fiducial marker system was constructed which attaches to a dental biteblock and has fidu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2007