Colvard pupillometer measurement of scotopic pupil diameter in emmetropes and myopes.

نویسندگان

  • Winai Chaidaroon
  • Wimolwan Juwattanasomran
چکیده

PURPOSE To prospectively compare the scotopic pupil size between emmetropes and myopes using a Colvard pupillometer. METHODS The pupil diameters of 55 normal subjects and 55 healthy myopic subjects were measured with the Colvard pupillometer in a low-light situation that simulated the level of light encountered while driving at night. RESULTS The mean (+/- SD) age of the emmetropic subjects was 30.78 years +/- 10.03 (range, 18-54 years) and the mean (+/- SD) age of the myopic subjects was 27.35 years +/- 8.43 (range, 21-52 years). The mean (+/- SD) scotopic pupil diameter was 6.46 +/- 0.90 mm (range, 4.5-8.0 mm) in the emmetropic group and 6.98 +/- 0.67 mm (5.5-8.5 mm) in the myopic group. The unpaired Student t-test showed that the difference in the scotopic pupil diameter between emmetropes and myopes was statistically significant (P =.0001). CONCLUSIONS The mean scotopic pupil diameter in myopes was larger than that in emmetropes. Therefore, a large ablation zone of the cornea or an appropriate optical size of the phakic intraocular lens should be considered in refractive surgery. Preoperative scotopic pupil measurements may be necessary in all refractive patients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of measurement of pupil sizes among the colvard pupillometer, procyon pupillometer, and NIDEK OPD-scan.

PURPOSE To compare pupil sizes measured with the Colvard pupillometer, Procyon pupillometer, and NIDEK OPD-Scan. METHODS Pupil diameter was measured in 90 consecutive eyes from 55 patients under mesopic and scotopic light conditions with all three instruments. RESULTS The mean scotopic pupil diameter was 6.3+/-0.98 mm with the Colvard pupillometer and 6.45+/-1.01 mm with the Procyon pupillo...

متن کامل

Pre-excimer laser and post-excimer laser refractive surgery measurements of scotopic pupil diameter using 2 pupillometers.

PURPOSE To compare a digital infrared pupillometer with a handheld light amplification pupillometer for measuring scotopic pupil size and to evaluate if the postoperative refractive changes of the cornea can influence pupil measurements. DESIGN Prospective noncomparative interventional case series. PARTICIPANTS One hundred eyes, 50 myopic (mean spherical equivalent [SE] refraction [+/- stan...

متن کامل

Evaluation of pupil dynamics after implantation of artisan phakic intraocular lenses.

PURPOSE To compare pre- and postoperative horizontal and vertical pupil diameters after horizontal Artisan phakic intraocular lens (PIOL) (Ophtec BV, Groningen, The Netherlands) implantation for correction of myopia. METHODS The pre- and postoperative scotopic pupil diameters measured by the Colvard pupillometer (Oasis Medical, Glendora, Calif) were compared in 71 eyes after Artisan PIOL impl...

متن کامل

Comparison of the Colvard, Procyon, and Neuroptics pupillometers for measuring pupil diameter under low ambient illumination.

PURPOSE To compare three different pupillometers (Colvard, Procyon, and Neuroptics) for determining pupil diameter at 0.04 and 0.4 lux ambient illumination. METHODS In 92 eyes of 46 healthy volunteers, pupil diameter was measured at 0.04 and 0.4 lux. After dark adaptation for 2 minutes, measurements were performed with each device by two examiners. Interobserver agreement, instrument agreemen...

متن کامل

Change in pupil size after implantation of an iris-fixated toric phakic intraocular lens.

PURPOSE To evaluate the changes in pupil size after implantation of an iris-supported toric phakic intraocular lens (TPIOL) for correction of myopia and hyperopia with astigmatism. SETTING Department of Ophthalmology, Johannes Gutenberg-University, Mainz, Germany. METHODS Twenty-two myopic eyes and 9 hyperopic eyes were included in the study. The mean age of the 2 groups was 34 years and 40...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Japanese journal of ophthalmology

دوره 46 6  شماره 

صفحات  -

تاریخ انتشار 2002