Comparison of temperature measurements and fundus reflectometry in laser coagulation.
نویسندگان
چکیده
In light coagulation, thermal Parameters play an important role for the biological reaction. Therefore, we measured the temperature increase and the whitening of the coagulation site simultaneously during the coagulation. Microthermocouples [I] were developed as shown in figure 1. The diameter of the tip ranges between 5 and 10 pm, the response time of this element is in the region of 1 msec. We designed a technique to insert this thermocouple into the globe through a trepanation at the Pars plana (fig. 2). The tip is advanced by a step motor and can be positioned exactly by a micromanipulater anywhere within the posterior pole. Coagulation of 500 ,um in diameter were set in rabbit eyes with an argon laser coagulator using a slitlamp contact lens arrangement. The temperature in the center of the lesion at the interface between the PE and retina is monitored during and after the coagulation (fig. 3). To demonstrate our method, figure 4 shows a typical temperature course duiing a weak coagulation. The light reflected from the coagulation site is monitored simultaneously with the temperature. The details of this technique have been reported in a paper at the 1976 Gonin Meeting in Lausanne by Bivngrubev et al. [2]. The upper curve of figure 4 shows the temperature course as a function of time, the lower curve shows the corresponding course of the reflccted light. In this case, the temperature increases by about 35 "L, but no change of the reflected light can be observed. This indicates that the
منابع مشابه
Noninvasive temperature measurements during laser irradiation of the retina with optoacoustic techniques
In all laser treatments at the fundus of the eye the temperature increase is unknown. In order to optimize the treatment modalities, a noninvasive online temperature determination is preferable. Method: Applying laser pulses to the fundus, thermoelastic stress waves are emitted based on the thermal expansion of the heated tissue, mainly the retinal pigment epithelium (RPE). The amplitude of the...
متن کاملMeasurement and comparison of the temperature-dependence of refractive index of water and Plexiglas phantoms by interferometry method for their use in optical calorimetry
In radiation calorimetry by using laser beams and interferometry setups, variations induced by dose absorption in phantom can be precisely measured. Dose absorption and the induced temperature change result in refractive index variation of the material. In order to be able to measure the low amount of absorbed dose in the phantom, temperature dependence of refractive index of the material must ...
متن کاملNoninvasive optoacoustic temperature determination at the fundus of the eye during laser irradiation.
In all fundus laser treatments of the eye, the temperature increase is not exactly known. In order to optimize treatments, an online temperature determination is preferable. We investigated a noninvasive optoacoustic method to monitor the fundus temperature during pulsed laser irradiation. When laser pulses are applied to the fundus, thermoelastic pressure waves are emitted, due to thermal expa...
متن کاملMeasured and predicted laser-induced temperature rises in the rabbit fundus.
Retinal damage caused by high-intensity light is generally due to a temperature increase in the fundus. Until the present time, it has not been possible to accurately measure this temperature rise. A procedure is described and results presented to show the dynamic temperature changes in the fundus caused by laser light and to compare the experimental results with a published retinal temperature...
متن کاملAutomatic temperature controlled retinal photocoagulation.
Laser coagulation is a treatment method for many retinal diseases. Due to variations in fundus pigmentation and light scattering inside the eye globe, different lesion strengths are often achieved. The aim of this work is to realize an automatic feedback algorithm to generate desired lesion strengths by controlling the retinal temperature increase with the irradiation time. Optoacoustics afford...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Modern problems in ophthalmology
دوره 20 شماره
صفحات -
تاریخ انتشار 1979