FlexPen and KwikPen Prefilled Insulin Devices: A Laboratory Evaluation of Ergonomic and Injection Force Characteristics.

نویسندگان

  • Debra A Ignaut
  • Michael Opincar
  • Sheila Lenox
چکیده

BACKGROUND Advances in insulin pen technology continue to improve the usability of these devices for patients with diabetes. In this study, ergonomic features and injection force, as measured by glide force (GF) and glide force variability (GFV), were evaluated for the new Humalog Mix75/25 KwikPen (KwikPen) and compared with the NovoLog Mix 70/30 FlexPen (FlexPen). METHODS Fifty prefilled insulin pen devices (25 of each type) were measured for diameter at the cartridge holder and dose window, length and weight with cap attached, and thumb reach at 30 and 60 units. GF was also determined for 100 devices (50 of each type); GFV at 30 and 60 unit doses was calculated for the plateau portion of the force curve based on the minimum and maximum force measured in that portion of the curve. RESULTS While FlexPen was lighter in weight than KwikPen, and presented a slightly smaller diameter at the cartridge holder and dose window, KwikPen had a shorter overall pen length compared to FlexPen, with a shorter thumb reach at both the 30- and 60-unit dose settings. The maximum GF for KwikPen was less than FlexPen at both the 30-unit (3.42 vs 5.36 lb, p <0.0001) and 60-unit doses (3.61 vs 5.62, p <0.0001). KwikPen GFV was lower across both doses (mean difference: -0.46 lb at 30 units, -0.44 lb at 60 units; p <0.0001 for both). CONCLUSIONS While FlexPen was lighter with a slightly smaller cartridge holder and dose window diameter, KwikPen was shorter in length with less thumb reach than FlexPen. KwikPen also demonstrated lower GF and GFV, resulting in a smoother injection profile than FlexPen. These features of KwikPen's design and function may offer important advantages for the user during insulin administration.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Injection force of SoloSTAR® compared with other disposable insulin pen devices at constant volume flow rates.

BACKGROUND Injection force is a particularly important practical aspect of therapy for patients with diabetes, especially those who have dexterity problems. This laboratory-based study compared the injection force of the SoloSTAR® insulin pen (SoloSTAR; sanofi-aventis) versus other available disposable pens at injection speeds based on the delivered volume of insulin released at the needle. M...

متن کامل

Analysis of the injection force of solostar® compared with other disposable insulin pen devices at constant volume flow rates.

Insulin pen devices have greatly enhanced the portability and accessibility to insulin therapy for millions of people with diabetes. Comparison research data should be reviewed thoroughly. In this issue of Journal of Diabetes Science and Technology, the study presented by Thomas van der Burg is balanced in number of samples tested, same tensile meter, and identical units per second delivery rat...

متن کامل

An evaluation of prefilled insulin pens: a focus on the Next Generation FlexPen®

Insulin pen delivery systems are preferred by patients over the traditional vial and syringe method for insulin delivery because they are simple and easy to use, improve confidence in dosing insulin, and have less interference with activities and improved discretion with use. Insulin manufacturers have made numerous improvements to their first marketed pen devices and are now introducing their ...

متن کامل

Accuracy and Injection Force of the Gla-300 Injection Device Compared With Other Commercialized Disposable Insulin Pens

BACKGROUND To deliver insulin glargine 300 U/mL (Gla-300), the widely used SoloSTAR(®) pen has been modified to allow for accurate and precise delivery of required insulin units in one-third of the volume compared with insulin glargine 100 U/mL, while improving usability. Here we compare the accuracy and injection force of 3 disposable insulin pens: Gla-300 SoloSTAR(®), FlexPen(®), and KwikPen™...

متن کامل

Insulin pen needles: effects of extra-thin wall needle technology on preference, confidence, and other patient ratings.

BACKGROUND Pen needles (PNs) are essential for insulin injections using pen devices. PN characteristics affect patients' injection experience. OBJECTIVE The goal of this study was to evaluate the impact of a new extra-thin wall (XTW) PN versus usual PNs on overall patient preference, ease of injection, perceived time to complete the full dose, thumb button force to deliver the injection, and ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of diabetes science and technology

دوره 2 3  شماره 

صفحات  -

تاریخ انتشار 2008