A High-Precision Micropipette Sensor for Cellular-Level Real-Time Thermal Characterization
نویسندگان
چکیده
We report herein development of a novel glass micropipette thermal sensor fabricated in a cost-effective manner, which is capable of measuring steady thermal fluctuation at spatial resolution of ∼2 μm with an accuracy of ±0.01 °C. We produced and tested various micrometer-sized sensors, ranging from 2 μm to 30 μm. The sensor comprises unleaded low-melting-point solder alloy (Sn-based) as a core metal inside a pulled borosilicate glass pipette and a thin film of nickel coating outside, creating a thermocouple junction at the tip. The sensor was calibrated using a thermally insulated calibration chamber, the temperature of which can be controlled with an accuracy of ±0.01 °C, and the thermoelectric power (Seebeck coefficient) of the sensor was recorded from 8.46 to 8.86 μV/°C. We have demonstrated the capability of measuring temperatures at a cellular level by inserting our temperature sensor into the membrane of a live retinal pigment epithelium cell subjected to a laser beam with a focal spot of 6 μm. We measured transient temperature profiles and the maximum temperatures were in the range of 38-55 ± 0.5 °C.
منابع مشابه
An Ant-Colony Optimization Clustering Model for Cellular Automata Routing in Wireless Sensor Networks
High efficient routing is an important issue for the design of wireless sensor network (WSN) protocols to meet the severe hardware and resource constraints. This paper presents an inclusive evolutionary reinforcement method. The proposed approach is a combination of Cellular Automata (CA) and Ant Colony Optimization (ACO) techniques in order to create collision-free trajectories for every agent...
متن کاملThermal analysis of ultrathin, compliant sensors for characterization of the human skin
Recent work establishes that ultrathin, stretchable sensors can enable high precision thermal characterization of the skin, with capabilities for spatial mapping, in forms that avoid irritation, thermal or mechanical loads on natural behaviors, or motion artifacts. The results have potential for use in cardiovascular screening, skin hydration sensing, and local skin heating and thermal therapy....
متن کاملThe near-field scanning thermal microscope.
We report on the design, characterization, and performance of a near-field scanning thermal microscope capable to detect thermal heat currents mediated by evanescent thermal electromagnetic fields close to the surface of a sample. The instrument operates in ultrahigh vacuum and retains its scanning tunneling microscope functionality, so that its miniature, micropipette-based thermocouple sensor...
متن کاملLimit of Detection of HEV RNA for Diagnosis of HEV Infection
ABSTRACT Background and Objectives: Diagnosis of hepatitis E virus (HEV) infection could be missed in some cases if serological tests are used solely. Molecular characterization of HEV is essential for diagnosis of acute and chronic HEV infections, and evaluating the chronic HEV infection status in immunocompromised patients. T...
متن کاملThe 3-Omega Method for the Measurement of Fouling Thickness, the Liquid Flow Rate, and Surface Contact
The 3-omega method is conventionally used for the measurement of thermal conductivity in solid samples. The present work includes the experimental characterization and proof-of-concept measurements of sensor concepts, based on the 3-omega method. It is shown that this method can be used to measure fouling layers with a thickness of 10 to 400 μm, to conduct the measurement of flow rates with a h...
متن کامل