A novel patient-specific navigational template for cervical pedicle screw placement.
نویسندگان
چکیده
STUDY DESIGN.: Prospective trial. OBJECTIVE.: To develop and validate a novel, patient-specific navigational template for cervical pedicle placement. SUMMARY OF BACKGROUND DATA.: Owing to the narrow bony anatomy and the proximity to the vertebral artery and the spinal cord, cervical instrumentation procedures demand the need for a precise technique for screw placement. PATIENT.: Specific drill template with preplanned trajectory has been thought as a promising solution for cervical pedicle screw placement. METHODS.: Patients with cervical spinal pathology (n = 25) requiring instrumentation were recruited. Volumetric CT scan was performed on each desired cervical vertebra and a 3-dimensional reconstruction model was generated from the scan data. Using reverse engineering technique, the optimal screw size and orientation were determined and a drill template was designed with a surface that is the inverse of the posterior vertebral surface. The drill template and its corresponding vertebra were manufactured using rapid prototyping technique and tested for violations. The navigational template was sterilized and used intraoperatively to assist with the placement of cervical screws. In total, 88 screws were inserted into levels C2-C7 with 2 to 6 screw in each patient. After surgery, the positions of the pedicle screws were evaluated using CT scan and graded for validation. RESULTS.: This method showed its ability to customize the placement and the size of each screw based on the unique morphology of the cervical vertebra. In all the cases, it was relatively very easy to manually place the drill template on the lamina of the vertebral body during the surgery. The required time between fixation of the template to the lamina and insertion of the pedicle screws was about 80 seconds. Of the 88 screws, 71 screws had no deviation and 14 screws had deviation <2 mm, 1 screw had a deviation between 2 to 4 mm and there were no misplacements. Fluoroscopy was used only once for every patient after the insertion of all the pedicle screws. CONCLUSION.: The authors have developed a novel patient-specific navigational template for cervical pedicle screw placement with good applicability and high accuracy. This method significantly reduces the operation time and radiation exposure for the members of the surgical team. The potential use of such a navigational template to insert cervical pedicle screws is promising. This technique has been clinically validated to provide an accurate trajectory for pedicle screw placement in the cervical spine.
منابع مشابه
A Novel Patient-Specific Drill Guide Template for Pedicle Screw Insertion into the Subaxial Cervical Spine Utilizing Stereolithographic Modelling: An In Vitro Study
STUDY DESIGN Cadaveric study. PURPOSE The purpose of this study was to assess the accuracy and feasibility of cervical pedicle screw (CPS) insertion into the subaxial cervical spine placed using a patient-specific drill guide template constructed from a stereolithographic model. OVERVIEW OF LITERATURE CPS fixation is an invaluable tool for posterior cervical fixation because of its biomecha...
متن کاملDeviation analysis of atlantoaxial pedicle screws assisted by a drill template.
Although C1-C2 pedicle screw fixation provides an excellent fusion rate and rigid fixation, this technique has a potential risk. It is essential to develop an accurate screwing method to avoid this neurovascular injury. To develop and validate the accuracy of a novel navigational template for C1-C2 pedicle screw placement in cadaveric specimens, computed tomography scans with 1-mm-wide cuts wer...
متن کاملA medium invasiveness multi-level patient’s specific template for pedicle screw placement in the scoliosis surgery
BACKGROUND Several methods including free-hand technique, fluoroscopic guidance, image-guided navigation, computer-assisted surgery system, robotic platform and patient's specific templates are being used for pedicle screw placement. These methods have screw misplacements and are not always easy to be applied. Furthermore, it is necessary to expose completely a large portions of the spine in or...
متن کاملA Modified Personalized Image-Based Drill Guide Template for Atlantoaxial Pedicle Screw Placement: A Clinical Study
BACKGROUND Atlantoaxial posterior pedicle screw fixation has been widely used for treatment of atlantoaxial instability (AAI). However, precise and safe insertion of atlantoaxial pedicle screws remains challenging. This study presents a modified drill guide template based on a previous template for atlantoaxial pedicle screw placement. MATERIAL AND METHODS Our study included 54 patients (34 mal...
متن کاملIndividualized 3D printing navigation template for pedicle screw fixation in upper cervical spine
PURPOSE Pedicle screw fixation in the upper cervical spine is a difficult and high-risk procedure. The screw is difficult to place rapidly and accurately, and can lead to serious injury of spinal cord or vertebral artery. The aim of this study was to design an individualized 3D printing navigation template for pedicle screw fixation in the upper cervical spine. METHODS Using CT thin slices da...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Spine
دوره 34 26 شماره
صفحات -
تاریخ انتشار 2009