Effects of Nebulizer Position, Gas Flow, and CPAP on Aerosol Bronchodilator Delivery: An In Vitro Study.

نویسندگان

  • Lorenzo Ball
  • Yuda Sutherasan
  • Valentina Caratto
  • Elisa Sanguineti
  • Maria Marsili
  • Pasquale Raimondo
  • Maurizio Ferretti
  • Robert M Kacmarek
  • Paolo Pelosi
چکیده

BACKGROUND The aim of this study was to investigate the effects of different delivery circuit configurations, nebulizer positions, CPAP levels, and gas flow on the amount of aerosol bronchodilator delivered during simulated spontaneous breathing in an in vitro model. METHODS A pneumatic lung simulator was connected to 5 different circuits for aerosol delivery, 2 delivering CPAP through a high-flow generator tested at 30, 60, and 90 L/min supplementary flow and 5, 10, and 15 cm H2O CPAP and 3 with no CPAP: a T-piece configuration with one extremity closed with a cap, a T-piece configuration without cap and nebulizer positioned proximally, and a T-piece configuration without cap and nebulizer positioned distally. Albuterol was collected with a filter, and the percentage amount delivered was measured by infrared spectrophotometry. RESULTS Configurations with continuous high-flow CPAP delivered higher percentage amounts of albuterol compared with the configurations without CPAP (9.1 ± 6.0% vs 6.2 ± 2.8%, P = .03). Among configurations without CPAP, the best performance was obtained with a T-piece with one extremity closed with a cap. In CPAP configurations, the highest delivery (13.8 ± 4.4%) was obtained with the nebulizer placed proximal to the lung simulator, independent of flow. CPAP at 15 cm H2O resulted in the highest albuterol delivery (P = .02). CONCLUSIONS Based on our in vitro study, without CPAP, a T-piece with a cap at one extremity maximizes albuterol delivery. During high-flow CPAP, the nebulizer should always be placed proximal to the patient, after the T-piece, using the highest CPAP clinically indicated.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In vitro evaluation of radio-labeled aerosol delivery via a variable-flow infant CPAP system.

BACKGROUND Nasal CPAP is widely used in neonatal ICUs. Aerosolized medications such as inhaled steroids and β agonists are commonly administered in-line through nasal CPAP, especially to infants with bronchopulmonary dysplasia. We hypothesized that aerosol delivery to the lungs via variable-flow nasal CPAP in an in vitro model would be unreliable, and that the delivery would depend on the posit...

متن کامل

Comparison of HFNC, bubble CPAP and SiPAP on aerosol delivery in neonates: An in-vitro study.

UNLABELLED Aerosol drug delivery via high flow nasal cannula (HFNC), bubble continuous positive airway pressure (CPAP), and synchronized inspiratory positive airway pressure (SiPAP) has not been quantified in spontaneously breathing premature infants. OBJECTIVES The purpose of this study was to compare aerosol delivery via HFNC, bubble CPAP, and SiPAP in a model of a simulated spontaneously b...

متن کامل

Aerosol delivery and humidification with the Boussignac continuous positive airway pressure device.

BACKGROUND A simple method for effective bronchodilator aerosol delivery while administering continuing continuous positive airway pressure (CPAP) would be useful in patients with severe bronchial obstruction. OBJECTIVE To assess the effectiveness of bronchodilator aerosol delivery during CPAP generated by the Boussignac CPAP system and its optimal humidification system. METHODS First we a...

متن کامل

Configuration of Aerosol Delivery Devices and Their Placement: Let's Be Sure We're Paying Attention to the Literature.

In this issue of RESPIRATORY CARE, Ball and colleagues1 determined the optimal nebulizer configurations and positions for aerosol delivery in vitro with and without CPAP. They determined that with continuous high-flow CPAP, the greatest amount of albuterol delivered to a lung simulator (13.8 4.4%) was when the nebulizer was placed proximal to the lung simulator. Without CPAP, they determined th...

متن کامل

Aerosol kinetics and bronchodilator efficacy during continuous positive airway pressure delivered by face mask.

BACKGROUND Rates of fresh gas flow (FGF) commonly used when continuous positive airway pressure (CPAP) is delivered by face mask theoretically reduce the delivery and availability of therapeutic aerosols. As it may be hazardous for patients with acute respiratory failure to interrupt mask CPAP, the effects of CPAP on aerosol kinetics and bronchodilator efficacy were investigated. METHOD The e...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 61 3  شماره 

صفحات  -

تاریخ انتشار 2016