False-negative urine protein electrophoresis by semiautomated gel electrophoresis.
نویسندگان
چکیده
Urine is evaluated by electrophoretic methods both to gather information about the location and degree of damage within the nephron and to detect and quantify monoclonal free light chains (Bence Jones proteins). For many years, urine protein electrophoresis consisted of concentrating the urine (usually 50-fold) followed by manual application of the concentrated sample to a gel. Recently, semiautomated methods have become available that allow for a more efficient application of urine to the gels. The Sebia Hydragel 12 15/30 (Sebia, Inc.) method provides urine gel results with crisp separation of 12 as recommended by the guidelines for clinical and laboratory evaluation of patients with monoclonal gammopathies (1 ). However, in a recent urine sample, despite the presence of 2288 mg/24 h of total protein, the urine protein electrophoresis gel failed to demonstrate the proteinuria although the urine had been concentrated 50-fold (Fig. 1A). The technologist had observed sediment in the concentrated urine specimen. Because of the discrepancy between the total protein concentration and the appearance of the urine on electrophoresis, the electrophoresis on this sample was repeated on the next run. Before this run, the urine was centrifuged for 5 min in a fixed-head Sero-fuge (Clay Adams). Studies on urine samples have shown that this centrifugation has no effect on serum protein migration (data not shown). After centrifugation, the clear supernatant was placed in the applicator well. The second electrophoresis demonstrated a glomerular and tubular proteinuria with a suspicious broad -region band (Fig. 1B). The difference in the two patterns indicated that no sample had been applied to the gel during the first analysis. The Sebia semiautomated agarose gel electrophoresis method uses plastic applicators (product no. 4199; Sebia) to apply the urine to the gel (Fig. 1C). Samples are pipetted into a plastic well, the bottom of which is a microporous membrane (Sebia) that leads to a straight edge. The liquid is drawn by wick effect into the microporous membrane and then to its edge. The applicator touches the edge of the microporous membrane to the gel, which places a small amount of the urine at the point of origin for the electrophoresis. The manufacturer’s instructions advise that urine with a high salt content may cause deformation of the gel. The instructions also note that use of old, improperly stored urine samples, in which enzymatic degradation of proteins might occur, should be avoided. During customer training, the company instructs
منابع مشابه
False-negative serum protein electrophoresis in a sample with an IgM monoclonal protein by semiautomated gel electrophoresis.
To the Editor: In a recent report, Keren et al. (1 ) described a case in which the Sebia semiautomated electrophoresis method failed to demonstrate proteinuria in a 50-fold concentrated urine sample despite the presence of 2 g/24 h total protein. The reason for this failure was attributed to the presence of debris associated with cellular elements that interfered with the wicking of the urine i...
متن کاملComparison of the serum Protein fractions of the developing chickens embryo by the technique of Starch-gel Electrophoresis
متن کامل
A cluster of culture positive gonococcal infections but with false negative cppB gene based PCR.
OBJECTIVES To describe the prevalence and characteristics of isolates of Neisseria gonorrhoeae grown from urine samples that produced negative results with nucleic acid amplification assays (NAA) targeting the cppB gene. METHODS An initial cluster of culture positive, but cppB gene based NAA negative, gonococcal infections was recognised. Urine samples and suspensions of gonococci isolated ov...
متن کاملPriority in clinical chemistry: does it match international public health need?
electrophoresis with the Hydrasys analyzer (Sebia), using Hydragel 15 HR gels (Sebia), missed a large IgM (12 g/L) monoclonal protein. The protein was detected by capillary electrophoresis and by the Beckman Paragon agarose electrophoresis system. The failure of the Hydrasys system to detect this monoclonal protein was related to the way sample application is performed with this semiautomated s...
متن کاملSimplified sodium dodecyl sulfate-polyacrylamide gel electrophoresis of unconcentrated urine with enhanced resolution and detection sensitivity.
We applied a simple sodium dodecyl sulfate-polyacrylamide gel electrophoresis method to urine. The method, developed for serum protein analysis (Clin Chem 1984;30:475-9), has a high sample throughput and gives excellent resolution with unconcentrated urine. It clearly distinguishes and characterizes proteinuric urine (7.5 microL) by Coomassie Blue staining and gives complex silver-stained patte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 50 5 شماره
صفحات -
تاریخ انتشار 2004