Aestiva ventilation mode selector switch failures.

نویسندگان

  • Dietrich Gravenstein
  • Harshdeep Wilkhu
  • Edwin B Liem
  • Stuart Tilman
  • Samsun Lampotang
چکیده

We describe three cases of previously unreported failures of the Bag-Ventilator Switch in Aestiva/5 anesthesia machines (GE Healthcare/Datex-Ohmeda, Madison, WI). Each failure mode produced a large breathing-circuit leak. Examination of the switches revealed a cracked toggle actuator, residue build-up, and a cracked selector switch housing as causes for the failures. When a leak with no visible cause develops, consider advancing the mode selector switch fully to its mechanical limit or consider that the toggle actuator or its anchoring mechanism may have failed. These cases demonstrate that it is imperative to always be prepared to immediately use an alternate method for ventilation. Cases describing failure to ventilate due to sudden equipment malfunction underscore the need to always have functioning backup ventilation equipment available.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Failure Mode and Effects Analysis Using Generalized Mixture Operators

Failure mode and effects analysis (FMEA) is a method based on teamwork to identify potential failures and problems in a system, design, process and service in order to remove them. The important part of this method is determining the risk priorities of failure modes using the risk priority number (RPN). However, this traditional RPN method has several shortcomings. Therefore, in this paper we p...

متن کامل

Second Order Sliding Mode Observer-Based Control for Uncertain Nonlinear MEMS Optical Switch

This paper studies theuncertain nonlinear dynamics of a MEMS optical switch addressing electrical, mechanical and optical subsystems. Recently, MEMS optical switch has had significant merits in reliability, control voltage requirements and power consumption. However, an inherent weakness in designing control for such systems is unavailability of switch position information at all times due to t...

متن کامل

Effective mode selector for tunable terahertz wire lasers.

We demonstrate an effective mode selector design that enables a terahertz quantum cascade wire laser to have a robust single-mode operation at frequencies much lower than the gain peak. This is achieved by selectively guiding the undesired modes into a lossy session while keeping the desired lasing mode largely unperturbed. The large mode discrimination obtained by this mode selector is necessa...

متن کامل

Safety assessment of glycol recovery unit in a gas refinery by failure mode and effects analysis technique

Background: The reliability and safety of gas refineries are strongly associated with the reliability of other parts of the system. Defect and failure in one part of a system can cause total system breakdown or accident. The failure and damage in these equipment such as transformers and boilers results in not only the equipment damage and human injuries but also productivity reduction. The purp...

متن کامل

Population genetic structure of the blue-fronted Amazon (Amazona aestiva, Psittacidae: Aves) based on nuclear microsatellite loci: implications for conservation.

The blue-fronted Amazon (Amazona aestiva) is a widely distributed Neotropical parrot and one of the most captured parrots in nature to supply the illegal trade of wild animals. The objectives of the present study were to analyze the genetic structure of A. aestiva to identify management units and support conservation planning and to verified if A. aestiva populations have undergone a recent bot...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Anesthesia and analgesia

دوره 104 4  شماره 

صفحات  -

تاریخ انتشار 2007