Chemical and clinical evaluation of the random access analyzer "RA-1000".

نویسندگان

  • M K Schwartz
  • B E Statland
  • J Coughlin
  • C Eisen
  • M Fleisher
  • R Ito
  • J B Levine
  • R Ng
  • G M Pennachia
  • H N Rose
چکیده

In the RA-1000, a random-access discrete analyzer, an inert fluorocarbon fluid is used to prevent interaction and carryover. Production-model instruments were evaluated in two laboratories with respect to determination of glucose, creatinine, total protein, inorganic phosphorus, cholesterol, alkaline phosphatase, lactate dehydrogenase, creatine kinase, gamma-glutamyltransferase, and aspartate and alanine aminotransferases. Within-run, among-run, and day-to-day (for 15 days) precision was assessed, and results were correlated with those obtained by the methods routinely in use in our departments. Precision was excellent, correlation acceptable.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An evaluation of the Technicon RA-1000 random-access analyzer.

We evaluated the Technicon RA-1000 analyzer, with emphasis on its potential for user-defined method development. Optical linearity and sample pipetting linearity were good. The reagent pipetting system delivered slightly less than the nominal amount, owing to the volumetric effect of added mixing bubbles. Carryover of aqueous solutions was negligible. The instrument had good adaptability for us...

متن کامل

The Technicon RA-1000 evaluated for measuring sodium, potassium, chloride, and carbon dioxide.

We evaluated the Technicon RA-1000 "random-access" analyzer for the measurements of sodium, potassium, and carbon dioxide by an indirect potentiometric method (ion-selective electrode) and for chloride by a colorimetric method (mercuric thiocyanate). For various concentrations of control materials the total precision (CV) ranged from 0.9 to 1.2% for sodium, 1.1 to 1.3% for potassium, 1.0 to 1.2...

متن کامل

Optimization and evaluation of cardiac enzymes and isoenzymes measured on a random access analyzer.

Four serum enzymes and isoenzymes used in the diagnosis of acute myocardial infarction (AMI), lactate dehydrogenase LD and LD-1, creatine kinase (CK), and CK-MB have been adapted to the Technicon RA-1000 automated clinical chemistry analyzer. Analytical parameters have been adjusted to provide clinically acceptable precision for all four assays. Correlations with centrifugal analyzer procedures...

متن کامل

Determinations of immunoglobulins G, A, and M in the Technicon RA-1000.

We evaluated the Technicon RA-1000 "Random Access" analyzer for the measurements of immunoglobulins G, A, and M in serum by turbidimetry. For various concentrations of serum pools the total CV ranged from 3.0 to 4.5% for IgG, from 3.1 to 3.2% for IgA, and from 4.8 to 5.4% for IgM. The linearity of the standard curves was good in the clinically useful ranges. Results for patients' samples correl...

متن کامل

Adaptation of EMIT drug assays to a random-access automated clinical analyzer.

Enzyme multiplied immunoassay technique (EMIT) assays for theophylline, carbamazepine, phenytoin, phenobarbital, valproic acid, acetaminophen, gentamicin and tobramycin, have been adapted to the Technicon RA-1000 random access analyzer. Working reagents are stable for a period of at least one week, and calibration need only be performed when preparing fresh reagents. Between-day precision range...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 30 3  شماره 

صفحات  -

تاریخ انتشار 1984