Oronasal and Tracheostomy Delivery of Soft Mist and Pressurized Metered-Dose Inhalers With Valved Holding Chamber.

نویسندگان

  • Ariel Berlinski
  • Brandy Cooper
چکیده

BACKGROUND Some pediatric tracheostomized patients who receive inhaled drugs undergo decannulation, and it is unknown whether the dose has to be adjusted. Pressurized metered-dose inhalers (pMDIs) and soft mist inhalers (SMIs) used with valved holding chambers (VHCs) made of non-electrostatic material are available. We hypothesized that using an SMI and changing the delivery route from tracheostomy to oronasal would increase lung dose. METHODS Four units of a metallic VHC were studied with albuterol hydrofluoroalkane (pMDI) and albuterol/ipratropium bromide with an SMI using an anatomically correct in vitro model of a 5-y-old spontaneously breathing tracheostomized child. The drug was captured in a filter and was termed lung dose. We tested breathing patterns with tidal volumes of 50, 155, and 300 mL. A mask and a special adapter were used as interfaces for oronasal and tracheostomy delivery, respectively. Spectrophotometry (276 nm) was used to determine albuterol concentration. RESULTS The use of SMI resulted in a higher lung dose than the pMDI for all tested conditions except delivery through tracheostomy with tidal volume of 155 mL (P = .69). Switching from oronasal to tracheostomy delivery increased the lung dose for all tested conditions except for the pMDI with the 300-mL tidal volume (P = .83). The use of SMI resulted in higher deposition in the tracheostomy tube than the pMDI. CONCLUSIONS In general, an SMI delivers a higher lung dose than a pMDI when using a metallic spacer during oronasal and tracheostomy route with the latter providing a higher lung dose.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics.

BACKGROUND Albuterol hydrofluoroalkane (HFA) either alone or with a valved holding chamber is used to treat bronchoconstriction. Delays between actuation and inhalation are common. Currently, the recommended interval between actuations of an albuterol HFA is 60 s. Using a shorter interval when multiple actuations are ordered will improve productivity in the hospital setting. METHODS We studie...

متن کامل

Force-dependent static dead space of face masks used with holding chambers.

BACKGROUND Pressurized metered-dose inhalers with valved holding chambers and masks are commonly used for aerosol delivery in children. Drug delivery can decrease when the dead-space volume (DSV) of the valved holding chamber is increased, but there are no published data evaluating force-dependent DSV among different masks. METHODS Seven masks were studied. Masks were sealed at the valved hol...

متن کامل

Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pMDI)

Valved holding chambers (VHCs) are medical devices that are recommended by International Asthma and Chronic Obstructive Pulmonary Disease (COPD) Guidelines [1, 2] to be used with pressurized metered dose inhalers (pMDI) as a delivery system. VHCs are designed to improve medication delivery, reduce oropharyngeal deposition of medication by substantially changing the aerodynamic particle size dis...

متن کامل

Jet nebulizers versus pressurized metered dose inhalers with valved holding chambers: effects of the facemask on aerosol delivery.

The delivery of an aerosolized drug to a child is a complex process requiring an interaction between parent, child, and inhalation device. Recent studies have shown that the facemask can be a key factor affecting aerosol delivery, particularly the influence of leaks between the facemask and the face. To further quantify these effects and design around them, we have developed a bench model consi...

متن کامل

The Bioavailability of Salbutamol in Urine via Volumatic and Nonvolumatic Valved Holding Chambers

PURPOSE Pressurized metered dose inhalers are commonly used in patients with asthma. However, the need to coordinate inhalation with inhaler actuation means that they are not suitable for use per se. Valved holding chamber devices were developed to overcome some of the problems of pressurized metered dose inhalers. Several types of holding chambers of different sizes are available in Iran. This...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 61 7  شماره 

صفحات  -

تاریخ انتشار 2016