Reducing Radiation Dose in Emergency CT Scans While Maintaining Equal Image Quality: Just a Promise or Reality for Severely Injured Patients?

نویسندگان

  • Ulrich Grupp
  • Max-Ludwig Schäfer
  • Henning Meyer
  • Alexander Lembcke
  • Alexander Pöllinger
  • Gero Wieners
  • Diane Renz
  • Philipp Schwabe
  • Florian Streitparth
چکیده

Objective. This study aims to assess the impact of adaptive statistical iterative reconstruction (ASIR) on CT imaging quality, diagnostic interpretability, and radiation dose reduction for a proven CT acquisition protocol for total body trauma. Methods. 18 patients with multiple trauma (ISS ≥ 16) were examined either with a routine protocol (n = 6), 30% (n = 6), or 40% (n = 6) of iterative reconstruction (IR) modification in the raw data domain of the routine protocol (140 kV, collimation: 40, noise index: 15). Study groups were matched by scan range and maximal abdominal diameter. Image noise was quantitatively measured. Image contrast, image noise, and overall interpretability were evaluated by two experienced and blinded readers. The amount of radiation dose reductions was evaluated. Results. No statistically significant differences between routine and IR protocols regarding image noise, contrast, and interpretability were present. Mean effective dose for the routine protocol was 25.3 ± 2.9 mSv, 19.7 ± 5.8 mSv for the IR 30, and 17.5 ± 4.2 mSv for the IR 40 protocol, that is, 22.1% effective dose reduction for IR 30 (P = 0.093) and 30.8% effective dose reduction for IR 40 (P = 0.0203). Conclusions. IR does not reduce study interpretability in total body trauma protocols while providing a significant reduction in effective radiation dose.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Ultra-Low Dose CT Image Reconstruction Based on Big Data Priors

1 Executive summary This project will develop a dramatically improved approach to low-dose X-ray CT image formation by extracting and using information from a big-data corpus of regular dose CT images. X-Ray Computed Tomography (CT) provides high-resolution images of anatomical structures for diagnosis and management of human diseases and disorders. For example, CT has had a tremendous impact o...

متن کامل

Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality

Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. Material and Methods: A poly(me...

متن کامل

Evaluation of Factors Affecting Absorbed Dose, Optimization Methods and Patient Dose Reduction in Dental Cone Beam CT (CBCT)

Introduction: Nowadays the use of cone beam computed tomography in dental imaging is increasing, although this method has a much lower dose than conventional CT scans, it delivers a higher dose than the panoramic and periapical patients therefore, , the aim of this study was to investigate the factors affecting the patients' dose in dental CBCTs and methods of optimizing and reducing the patien...

متن کامل

Image Optimization and reduction of radiation dose in CT of the paranasal sinuses

Introduction: Since demand for computed tomography (CT) examinations is growing, patients are exposed more frequently to ionizing radiation. To reduce the dose delivered to patients in each scan, CT technologists can change the image acquisition parameters. However, this reduces image quality. Present study aimed to optimize and reduce radiation dose in a CT of the paranasal si...

متن کامل

Prospective-triggered high-pitch spiral versus sequential dual-source CT coronary angiography: comparison of image quality and radiation dose

Background: Prospectively electrocardiography (ECG)-triggered high-pitch spiral coronary computed tomography angiography (CCTA) is a unique scan mode for dual-source CT (DSCT). Our reports aim to compare image quality and radiation dose of CCTA using high-pitch spiral or sequential acquisition mode in patients with low and stable heart rates. Materials and Methods: Patients with low and stable ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2013  شماره 

صفحات  -

تاریخ انتشار 2013