Cardiac Output Monitoring: Validation Studies–how Results Should be Presented

نویسندگان

  • Peter M. Odor
  • Sohail Bampoe
  • Maurizio Cecconi
چکیده

Purpose of Review Cardiac output monitors can be assessed by a variety of techniques, but a common principle is quantifying agreement between a reference standard and new monitor. The current standard analysis technique is a Bland-Altman plot. The Bland-Altman plot evaluates bias between mean differences of cardiac output, from which an agreement interval is derived. These limits are, however, statistical limits of agreement and the clinical acceptability will depend upon context and application. This article provides suggestions for understanding and presenting the results of cardiac output validation, using standard metrology alongside proposals for criteria used to accept new techniques. Recent Findings Confusion about the appropriate way to report "precision" in method comparison studies stem from a lack of clarity on how single or repeated measurements should be interpreted. During serial measurements of cardiac output the true value changes, thus measurement should be considered as serial rather than repeated. Method agreement based upon precision achieved by cardiac output monitors needs to consider each method's general variability around true values obtained and this data should be generated and presented as part of each study design. Summary Studies should report serial measurements from two techniques for cardiac output monitoring. Results of similar techniques from other studies may not always be transferred and compared. Bias and intervals of agreement should be presented as Bland-Altman plots with dynamic cardiac output trends in polar plots. Percentage error should be calculated to allow appropriate comparison of techniques for study populations with different expected cardiac output values.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Bias and precision statistics: should we still adhere to the 30% benchmark for cardiac output monitor validation studies?

To the Editor: It was with great interest that I read Peyton and Chong’s meta-analysis of papers comparing minimally invasive cardiac output monitors to thermodilution, particularly as much of the discussion centered around a much-cited paper by me and my late brother Julian Critchley. Our 1999 paper made a number of important contributions to assessing cardiac output monitoring: (1) it provide...

متن کامل

What technique should I use to measure cardiac output?

PURPOSE OF REVIEW Several less invasive cardiac output monitoring techniques are now commercially available and have the potential to replace the pulmonary artery catheter under certain clinical circumstances. The aim of this review is to give a synopsis of the currently available cardiac output measurement methods. This information should help in selecting the appropriate technique in a partic...

متن کامل

A System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis

Background: Cardiac output (CO) is the total volume of blood pumped by the heart per minute and is a function of heart rate and stroke volume. CO is one of the most important parameters for monitoring cardiac function, estimating global oxygen delivery and understanding the causes of high blood pressure. Hence, measuring CO has always been a matter of interest to researchers and clinicians. Sev...

متن کامل

Noninvasive continuous cardiac output monitoring in perioperative and intensive care medicine.

The determination of blood flow, i.e. cardiac output, is an integral part of haemodynamic monitoring. This is a review on noninvasive continuous cardiac output monitoring in perioperative and intensive care medicine. We present the underlying principles and validation data of the following technologies: thoracic electrical bioimpedance, thoracic bioreactance, vascular unloading technique, pulse...

متن کامل

Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output

Bland-Altman analysis is used for assessing agreement between two measurements of the same clinical variable. In the field of cardiac output monitoring, its results, in terms of bias and limits of agreement, are often difficult to interpret, leading clinicians to use a cutoff of 30% in the percentage error in order to decide whether a new technique may be considered a good alternative. This per...