Accuracy and precision of fourteen pulse oximeters.

نویسندگان

  • B Hannhart
  • J P Haberer
  • C Saunier
  • M C Laxenaire
چکیده

Two sets of seven pulse oximeters (Criticare CSI-502; Nellcor-N200; Datex-Satlite; Physio-Control-Lifestat 1600; Critikon-Oxyshuttle; Ohmeda-Biox 3700; Ohmeda-Biox 3740; Radiometer-Oxi; Spectramed-Pulsat; Kontron-7840; Biochem-Ox2000; Invivo-4500; Engström-EOS; Novametrix-505) were studied in two groups of eight healthy subjects, aged 26-50 yrs. The transcutaneous oxygen saturation (SpO2) was compared with arterial oxygen saturation (SaO2) measured in simultaneously with drawn blood samples (OSM2 Radiometer) at four 20 min steady-state levels of inspired oxygen fraction (FIO2) (0.21, 0.10, 0.08 and 0.07; SaO2 99-55%) in a conditioned chamber. Both the error in accuracy (mean SpO2-SaO2 difference), and the error in precision (SD of the differences) remained less than 3% for the two highest FIO2 levels (SaO2 greater than 83%) but, during deeper hypoxia, they were increased to 8% and 5%, respectively. An instrumental systematic bias affected accuracy in particular. We concluded that a good agreement between SpO2 and SaO2, as reflected by the Bartko's intraclass coefficient, was observed in nine instruments.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Bias and precision of pulse oximeters and arterial oximeters.

We evaluated measurements of oxygen saturation from four noninvasive pulse oximeters, and two conventional arterial oximeters. Simultaneous measurements were obtained on each instrument on three different occasions in five healthy subjects breathing gas with an FIO2 of 1.00, 0.50, 0.21, 0.17, 0.15, 0.13 and 0.11. Mean bias relative to the sum of oxy-, carboxy-, and methemoglobins ranged from -0...

متن کامل

Accuracy of Pulse Oximeters in Detecting Hypoxemia in Patients with Chronic Thromboembolic Pulmonary Hypertension

PURPOSE Pulse oximetry is routinely used to continuously and non-invasively monitor arterial oxygen saturation (SaO2). When oxygen saturation by pulse oximeter (SpO2) overestimates SaO2, hypoxemia may be overlooked. We compared the SpO2 - SaO2 differences among three pulse oximeters in patients with chronic thromboembolic pulmonary hypertension (CTEPH) who spent their daily lives in a poor oxyg...

متن کامل

Variability in estimating shunt from single pulse oximetry measurements.

Virtual shunt describes the overall loss of O2 content between the alveolar gas and arterial blood. Clinicians indirectly estimate the magnitude of the virtual shunt by monitoring peripheral blood oxygen saturation (SpO2) using non-invasive pulse oximetry. An inherent limitation of this method is the variable precision of pulse oximeters and the non-linear relationship between virtual shunt and...

متن کامل

Pulse oximetry in children with congenital heart disease: effects of cardiopulmonary bypass and cyanosis.

The objective of this prospective, observational study with consecutive sampling was to assess the reliability, bias, and precision of Nellcor N-395 (N) and Masimo SET Radical (M) pulse oximeters in children with cyanotic congenital heart disease and children with congenital heart disease recovering from cardiopulmonary bypass-assisted surgery admitted to a cardiovascular operating suite and pe...

متن کامل

Accuracy of pulse oximeters in estimating heart rate at rest and during exercise.

Pulse oximeters are being widely used for non-invasive, simultaneous assessment of haemoglobin oxygen saturation. They are reliable, accurate, relatively inexpensive and portable. Pulse oximeters are often used for estimating heart rate at rest and during exercise. However, at present the data available to validate their use as heart rate monitors are not sufficient. We evaluated the accuracy o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 4 1  شماره 

صفحات  -

تاریخ انتشار 1991