Echocardiographic guidance during placement of the buttoned double-disk device for atrial septal defect closure.

نویسندگان

  • L L Minich
  • A R Snider
چکیده

The usefulness of two-dimensional and Doppler echocardiography during buttoned double-disk device closure of an atrial septal defect was evaluated in 20 consecutive patients at the time of interventional catheterization. Transesophageal echocardiography was used in 11 patients (ages 5 to 62 years, weights 20 to 91 kg). Because of the size of the available transesophageal echo probe, transthoracic echocardiography was used in the remaining 9 patients (ages 4 to 5.5 years, weights 14 to 21 kg). In the transesophageal echo group, 1 patient was found to have no atrial septal defect despite a previous diagnosis by transthoracic echocardiography, 3 patients had atrial septal defects too large for closure despite attempts in 2, and 7 patients had transesophageal echo guided device placement. All of these 7 patients had small residual shunts by color Doppler, 2 had unusual arm positions, and 2 had surgical removal of the device due to embolization to the pulmonary artery in 1 and failure to obtain close approximation of the occluder and counteroccluder in 1. In the transthoracic echo group, 2 patients had atrial septal defects too large for closure, 1 patient had no femoral venous access, and 6 patients had transthoracic echo guided device placement. All of these 6 patients had small residual shunts by color Doppler and 3 of the 6 had unusual arm positions. For atrial septal defect sizing, transesophageal echo measurements correlated with catheter balloon size more closely than did transthoracic echo measurements (r 2 = 0.97 vs 0.86).(ABSTRACT TRUNCATED AT 250 WORDS)

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Influence of Echocardiographic Guidance on Positioning of the Buttoned Occluder for Transcatheter Closure of Atrial Septal Defects.

Ideal position of the buttoned device for occlusion of atrial septal defects was achieved more often with transesophageal than with transthoracic echocardiographic guidance (10/11 vs 7/23; P = 0.001). Patients with ideal device position were less likely to have residual shunts, device unbuttoning, or atrioventricular valve regurgitation (2/17 vs 11/17; P = 0.002). We therefore recommend the use...

متن کامل

Transcatheter closure of residual atrial septal defect following implantation of buttoned device.

We report a case in which residual shunting after a buttoned device occlusion of atrial septal defect (ASD) was eliminated by transcatheter retrieval of a portion of the device, followed by implantation of a second device. This method may be helpful for those patients with residual ASDs who decline surgical device retrieval and defect closure.

متن کامل

Urgent Surgical Intervention for Embolized Cardiac Occluder Devices: A Case Series

Introduction: In this study, we sought to illustrate our experience in urgent surgical management for embolized cardiac septal occlude devices resulting from trans-catheter closure of atrial septal defect and ventricular septal defect. Mathrials and Methods: We retrospectively reviewed four patients aged 2–10 years who underwent urgent surgery due to cardiac septal occluder embolization between...

متن کامل

Peratrial Device Closure of Atrial Septal Defect Under Transesophageal Echocardiographic Guidance without Fluoroscopy Compared to Conventional On-Pump Surgical Closure

Objective This study is designed to evaluate the advantages between peratrial device closure under transesophageal echocardiographic guidance and open heart surgery in atrial septal defect. Methods From November 2011 to September 2014, 28 patients with atrial septal defect were treated. Fourteen patients received peratrial device closure under transesophageal echocardiographic guidance (TEE g...

متن کامل

Transvenous atrial septal defect occlusion in piglets with a "buttoned" double-disk device.

The feasibility and safety of transvenous closure of atrial septal defects by a new device was tested in 20 piglets, in which atrial septal defects were created by foramen ovale dilatation with angioplasty balloons. The device was small enough to be introduced in a 7F or 8F sheath, and it measured 20-25 mm. It has no hooks and consists of a foam occluder from the left atrium and a counter-occlu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Echocardiography

دوره 10 6  شماره 

صفحات  -

تاریخ انتشار 1993