Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
نویسندگان
چکیده
BACKGROUND This study analyzes the characteristics of donor and recipient tissue preparation between the Hessburg-Barron and Hanna punch and trephine systems by using elliptical curve fitting models, light microscopy, and anterior segment optical coherence tomography (AS-OCT). METHODS Eight millimeter Hessburg-Barron and Hanna vacuum trephines and punches were used on six cadaver globes and six corneal-scleral rims, respectively. Eccentricity data were generated using measurements from photographs of the corneal buttons and were used to generate an elliptical curve fit to calculate properties of the corneal button. The trephination angle and punch angle were measured by digital protractor software from light microscopy and AS-OCT images to evaluate the consistency with which each device cuts the cornea. RESULTS The Hanna trephine showed a trend towards producing a more circular recipient button than the Barron trephine (ratio of major axis to minor axis), ie, 1.059 ± 0.041 versus 1.110 ± 0.027 (P = 0.147) and the Hanna punch showed a trend towards producing a more circular donor cut than the Barron punch, ie, 1.021 ± 0.022 versus 1.046 ± 0.039 (P = 0.445). The Hanna trephine was demonstrated to have a more consistent trephination angle than the Barron trephine when assessing light microscopy images, ie, ±14.39° (95% confidence interval [CI] 111.9-157.7) versus ±19.38° (95% CI 101.9-150.2, P = 0.492) and OCT images, ie, ±8.08° (95% CI 106.2-123.3) versus ±11.16° (95% CI 109.3-132.6, P = 0.306). The angle created by the Hanna punch had less variability than the Barron punch from both the light microscopy, ie, ±4.81° (95% CI 101.6-113.9) versus ±11.28° (95% CI 84.5-120.6, P = 0.295) and AS-OCT imaging, ie, ±9.96° (95% CI 95.7-116.4) versus ±14.02° (95% CI 91.8-123.7, P = 0.825). Statistical significance was not achieved. CONCLUSION The Hanna trephine and punch may be more accurate and consistent in cutting corneal buttons than the Hessburg-Barron trephine and punch when evaluated using elliptical curve fitting models, light microscopy, and AS-OCT.
منابع مشابه
Accuracy of corneal trephination depth using the Moria single-use adjustable depth vacuum trephine system
BACKGROUND The Moria single-use adjustable depth trephine is a device that allows a goal trephination depth to be set prior to the surgical procedure. METHODS Eleven fresh human cadaveric eyes were trephined using 8.0 mm Moria single-use adjustable vacuum trephines. Prior to trephination, the average corneal pachymetry in the peripheral 7-10 mm range was obtained using anterior segment optica...
متن کاملDevelopment of an Advanced Optical Coherence Tomography System for Radiation Dosimetry
Introduction: According to the literature, optical coherence tomography (OCT) can be used measure radiation absorbed dose. This study was carried out to design a computed tomography system for the calculation of absorbed dose and optimization of dose delivery in radiotherapy using gel dosimeters. Material and Methods: An advanced charge-coupled device based OCT system was developed in laborator...
متن کاملExperimental Visualization of Labyrinthine Structure with Optical Coherence Tomography
Introduction:Visualization of inner ear structures is a valuable strategy for researchers and clinicians working on hearing pathologies. Optical coherence tomography (OCT) is a high-resolution imaging technology which may be used for the visualization of tissues. In this experimental study we aimed to evaluate inner ear anatomy in well-prepared human labyrinthine bones.Materials and Methods:Thr...
متن کاملComparison of Choroidal Thickness in Psoriasis and Control Groups Using Optical Coherence Tomography
Background and Objective: One of the most common extracutaneous complications in psoriasis patients is ocular complications, which is controversial concerning its effect on choroidal thickness. This study aimed to compare the psoriasis patients with healthy individuals in terms of the thickness of the choroidal layer using the optical coherence tomography technique. Materials and Methods: In t...
متن کاملCorrelation of anterior segment optical coherence tomography measurements with graft trephine diameter following descemet stripping automated endothelial keratoplasty
BACKGROUND To assess repeatability of the Zhongshan Assessment Program (ZAP) software measurement of Anterior Segment Optical Coherence Tomography (ASOCT) images and correlate with graft trephine diameter following Descemet Stripping Automated Endothelial Keratoplasty (DSAEK) METHODS: Retrospectively evaluated interventional case series. 121 consecutive eyes undergoing DSAEK over a 26 month per...
متن کامل