Fully automated, internally controlled quantification of hepatitis B Virus DNA by real-time PCR by use of the MagNA Pure LC and LightCycler instruments.

نویسندگان

  • Victoria Leb
  • Markus Stöcher
  • Elizabeth Valentine-Thon
  • Gabriele Hölzl
  • Harald Kessler
  • Herbert Stekel
  • Jörg Berg
چکیده

We report on the development of a fully automated real-time PCR assay for the quantitative detection of hepatitis B virus (HBV) DNA in plasma with EDTA (EDTA plasma). The MagNA Pure LC instrument was used for automated DNA purification and automated preparation of PCR mixtures. Real-time PCR was performed on the LightCycler instrument. An internal amplification control was devised as a PCR competitor and was introduced into the assay at the stage of DNA purification to permit monitoring for sample adequacy. The detection limit of the assay was found to be 200 HBV DNA copies/ml, with a linear dynamic range of 8 orders of magnitude. When samples from the European Union Quality Control Concerted Action HBV Proficiency Panel 1999 were examined, the results were found to be in acceptable agreement with the HBV DNA concentrations of the panel members. In a clinical laboratory evaluation of 123 EDTA plasma samples, a significant correlation was found with the results obtained by the Roche HBV Monitor test on the Cobas Amplicor analyzer within the dynamic range of that system. In conclusion, the newly developed assay has a markedly reduced hands-on time, permits monitoring for sample adequacy, and is suitable for the quantitative detection of HBV DNA in plasma in a routine clinical laboratory.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Prevalence of Occult Hepatitis B Infection Among Blood Donors with HBsAg Negative and HBcAb Positive by Real Time PCR in South of Iran

Abstract   Background & aim: Occult hepatitis B virus infection in blood donors threats the safety of the blood supply.  It is characterized by detection of low viral DNA in the serum subjects who have negative test for Hepatitis B surface antigen (HBsAg). The aim of this study was to assess the prevalence of occult hepatitis B virus infection among HBsAg negative and HBcAb positive in blood ...

متن کامل

ارزیابی تکنیک های ICT،ELISA و Real Time PCR در تشخیص هپاتیت B در بیماران دیالیزی،تالاسمی و هموفیلی

Background and purpose: Recently, use of enzyme-linked immunosorbent assay (ELISA) and rapid Immunochromatographic test (ICT) in diagnosis and screening of patients with hepatitis B reduced the risk of hepatitis during blood transfusion. However, the incidence of hepatitis B is very high in high-risk patients such as hemodialysis, thalassemia and hemophilia which receive a lot of blood transfus...

متن کامل

Use of the MagNA pure LC automated nucleic acid extraction system followed by real-time reverse transcription-PCR for ultrasensitive quantitation of hepatitis C virus RNA.

Hepatitis C virus (HCV) infection is an increasing health problem worldwide. Quantitative assays for HCV viral load are valuable in predicting response to therapy and for following treatment efficacy. Unfortunately, most quantitative tests for HCV RNA are limited by poor sensitivity. We have developed a convenient, highly sensitive real-time reverse transcription-PCR assay for HCV RNA. The assa...

متن کامل

کلونینگ ژن کدکننده آنتی‌ژن سطحی ویروس هپاتیت B در اشرشیاکولی

    Background & Aim: Hepatitis B virus(HBV) infection is endemic worldwide. It is estimated every year more than 350 million people become infected with HBV(new cases) worldwide. Unfortunately, there are no satisfactory drugs to cure HBV and related diseases and the only way to control it is through vaccination. Measurements of HBV DNA levels are routinely used to identify infectious chronic c...

متن کامل

Real-time PCR coupled with automated DNA extraction and detection of galactomannan antigen in serum by enzyme-linked immunosorbent assay for diagnosis of invasive aspergillosis.

To improve the diagnosis of invasive aspergillosis (IA), we developed a LightCycler PCR assay targeted to Aspergillus fumigatus and A. flavus mitochondrial DNA. To avoid contamination, fully automated nucleic acid extraction with the MagNA Pure LC apparatus was used. The linearity of the results was achieved over a 6-log range of input A. fumigatus DNA, from 0.3 ng to 3 fg/10 microl of water. W...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 42 2  شماره 

صفحات  -

تاریخ انتشار 2004