Improved setup and positioning accuracy using a three‐point customized cushion/mask/bite‐block immobilization system for stereotactic reirradiation of head and neck cancer

نویسندگان

  • He Wang
  • Congjun Wang
  • Samuel Tung
  • Andrew Wilson Dimmitt
  • Pei Fong Wong
  • Mark A. Edson
  • Adam S. Garden
  • David I. Rosenthal
  • Clifton D. Fuller
  • Gary B. Gunn
  • Vinita Takiar
  • Xin A. Wang
  • Dershan Luo
  • James N. Yang
  • Jennifer Wong
  • Jack Phan
چکیده

The purpose of this study was to investigate the setup and positioning uncertainty of a custom cushion/mask/bite-block (CMB) immobilization system and determine PTV margin for image-guided head and neck stereotactic ablative radiotherapy (HN-SABR). We analyzed 105 treatment sessions among 21 patients treated with HN-SABR for recurrent head and neck cancers using a custom CMB immobilization system. Initial patient setup was performed using the ExacTrac infrared (IR) tracking system and initial setup errors were based on comparison of ExacTrac IR tracking system to corrected online ExacTrac X-rays images registered to treatment plans. Residual setup errors were determined using repeat verification X-ray. The online ExacTrac corrections were compared to cone-beam CT (CBCT) before treatment to assess agreement. Intrafractional positioning errors were determined using prebeam X-rays. The systematic and random errors were analyzed. The initial translational setup errors were -0.8 ± 1.3 mm, -0.8 ± 1.6 mm, and 0.3 ± 1.9 mm in AP, CC, and LR directions, respectively, with a three-dimensional (3D) vector of 2.7 ± 1.4 mm. The initial rotational errors were up to 2.4° if 6D couch is not available. CBCT agreed with ExacTrac X-ray images to within 2 mm and 2.5°. The intrafractional uncertainties were 0.1 ± 0.6 mm, 0.1 ± 0.6 mm, and 0.2 ± 0.5 mm in AP, CC, and LR directions, respectively, and 0.0° ± 0.5°, 0.0° ± 0.6°, and -0.1° ± 0.4° in yaw, roll, and pitch direction, respectively. The translational vector was 0.9 ± 0.6 mm. The calculated PTV margins mPTV(90,95) were within 1.6 mm when using image guidance for online setup correction. The use of image guidance for online setup correction, in combination with our customized CMB device, highly restricted target motion during treatments and provided robust immobilization to ensure minimum dose of 95% to target volume with 2.0 mm PTV margin for HN-SABR.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Six-Dimensional Cranial Target Positioning Accuracy in Two Different Immobilization Methods Using Exactrac System

Introduction: The aim of this study was to determine the accuracy of two different immobilization methods in patient positioning in cranial radiotherapy. The six-dimensional (6D) target localization accuracy of using a dedicated stereotactic mask was compared with that of a conventional head mask by the ExacTrac system. Material and Methods: ...

متن کامل

Spine Stereotactic Body Radiation Therapy Residual Setup Errors and Intra-Fraction Motion Using the Stereotactic X-Ray Image Guidance Verification System

Purpose To determine the precision of our institution's current immobilization devices for spine SBRT, ultimately leading to recommendations for appropriate planning margins. Methods We identified 12 patients (25 treatments) with spinal metastasis treated with spine Stereotactic Body Radiation Therapy (SBRT). The Body-FIX system was used as immobilization device for thoracic (T) and lumbar (L...

متن کامل

Evaluation of the setup accuracy of a stereotactic radiotherapy head immobilization mask system using kV on‐board imaging†

The purpose of this study was to evaluate setup accuracy and quantify random and systematic errors of the BrainLAB stereotactic immobilization mask and localization system using kV on-board imaging. Nine patients were simulated and set up with the BrainLAB stereotactic head immobilization mask and localizer to be treated for brain lesions using single and hypofractions. Orthogonal pairs of proj...

متن کامل

A novel setup approach for helical tomotherapy in head and neck cancer: A case report

Head and neck cancers that are associated with high-risk factors have a poor prognosis. In such patients, post-operative radiochemotherapy is mandatory. The present study describes the case of a 71-year-old high-risk head and neck cancer patient who was not able to tolerate the supine position for the radiotherapy setup. A lateral immobilization with a head mask and a vacuum cushion was perform...

متن کامل

Stereotactic IMRT for prostate cancer: Setup accuracy of a new stereotactic body localization system

The purpose of this work is to prospectively assess the setup accuracy that can be achieved with a stereotactic body localizer (SBL) in immobilizing patients for stereotactic intensity-modulated radiotherapy (IMRT) for prostate cancer. By quantifying this important factor and target mobility in the SBL, we expect to provide a guideline for selecting planning target volume margins for stereotact...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016