A comparison of the Airtraq®, McGrath®, and Macintosh laryngoscopes for difficult paediatric intubation: A manikin study
نویسندگان
چکیده
BACKGROUND The efficacy of devices for difficult intubation in paediatric patients, especially with a Cormack-Lehane grade 4 view, has yet to be established. We compared intubating parameters among three devices (the Airtraq®, McGrath®, and Macintosh laryngoscopes). METHODS This study is a randomised cross-over trial. Participants were 20 anaesthetists. Each device was tested three times using a paediatric manikin with a Cormack-Lehane grade 4 view. The order to use each device was randomised by a computer-generated random sequence. The primary endpoint was the rate of successful intubation. Secondary endpoints included the time taken to intubate, percentage of glottic opening score, and severity of potential dental trauma. RESULTS The successful intubation rates of the Airtraq®, McGrath®, and Macintosh laryngoscopes were 100%, 72%, and 45%, respectively. The risk ratio of the success rates of Airtraq® compared with McGrath® and Macintosh laryngoscopes were 1.40 (95% CI; 1.19-1.64, P < 0.001) and 2.22 (95% CI; 1.68-2.94, P < 0.001), respectively. The modified Cormack-Lehane grade and percentage of the glottic opening score were better for the Airtraq® than for the other devices. The dental trauma score was lower for the Airtraq® than for the other devices. There were no significant differences in the intubation time among the groups. CONCLUSIONS The Airtraq® had higher success rate, had better visibility, and was associated with less dental trauma than the other devices in a difficult paediatric intubation model with a Cormack-Lehane grade 4 view.
منابع مشابه
A randomised crossover trial comparing the Airtraq(®) NT, McGrath(®) MAC and Macintosh laryngoscopes for nasotracheal intubation of simulated easy and difficult airways in a manikin.
OBJECTIVE Several devices can aid nasotracheal intubation when managing difficult airways. The McGrath MAC and Airtraq NT were compared with a Macintosh laryngoscope when studying the performance of anaesthetists with different levels of experience, in a manikin model of easy or difficult airway scenarios. METHODS Sixty-three anaesthetists were recruited into a randomised trial in which each ...
متن کاملComparison of the Airtraq® and Truview® laryngoscopes to the Macintosh laryngoscope for use by Advanced Paramedics in easy and simulated difficult intubation in manikins
BACKGROUND Paramedics are frequently required to perform tracheal intubation, a potentially life-saving manoeuvre in severely ill patients, in the prehospital setting. However, direct laryngoscopy is often more difficult in this environment, and failed tracheal intubation constitutes an important cause of morbidity. Novel indirect laryngoscopes, such as the Airtraq and Truview laryngoscopes may...
متن کاملComparing McGRATH® MAC, C-MAC®, and Macintosh Laryngoscopes Operated by Medical Students: A Randomized, Crossover, Manikin Study
We hypothesized that the McGRATH MAC would decrease the time of intubation compared to C-MAC for novices. Thirty-nine medical students who had used the Macintosh blade to intubate a manikin fewer than 3 times were recruited. The participants performed sequential intubations on the manikin in two simulated settings that included a normal airway and a difficult airway (tongue edema). The intubati...
متن کاملAirtraq, LMA CTrach and Macintosh Laryngoscopes in Tracheal Intubation Training: A Randomized Comparative Manikin Study.
OBJECTIVE Training students on simulators before allowing their direct contact with patients is well accepted. There is no clinical or manikin-based simulation study in the literature comparing tracheal intubation with Airtraq, laryngeal mask airway (LMA) CTrach and Macintosh laryngoscopes performed by medical students having no prior intubation experience. METHODS After obtaining written inf...
متن کاملEvaluation of intubation using the Airtraq or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy--a manikin study.
The Airtraq Laryngoscope is a novel intubation device which allows visualisation of the vocal cords without alignment of the oral, pharyngeal and tracheal axes. We compared the Airtraq with the Macintosh laryngoscope in simulated easy and difficult laryngoscopy. Twenty-five anaesthetists were allowed up to three attempts to intubate the trachea in each of three laryngoscopy scenarios using a La...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 12 شماره
صفحات -
تاریخ انتشار 2017