Forehead pulse oximetry: Headband use helps alleviate false low readings likely related to venous pulsation artifact.
نویسندگان
چکیده
BACKGROUND This study investigated whether a tensioning headband that applies up to 20 mmHg pressure over a forehead pulse oximetry sensor could improve arterial hemoglobin oxygen saturation reading accuracy in presence of venous pooling and pulsations at the forehead site. METHODS Healthy volunteers were studied breathing room air in supine and various levels of negative incline (Trendelenburg position) using the forehead sensor with the headband adjusted to its maximum and minimum recommended pressure limits. Saturation readings obtained from the forehead sensor with the subjects supine and the headband in place were used as a baseline to compare the effects of negative incline on reading accuracy when using and not using the headband. Occurrences of false low-saturation readings detected by forehead sensors were compared with those from digit sensors. RESULTS No difference was observed between saturation readings obtained from the forehead sensor in supine and negative incline positions when the headband was applied. Forehead sensor readings obtained while subjects were inclined and the headband was not used were significantly lower (P < 0.05) than the supine readings. There was no statistically significant difference between the digit and forehead sensor in reporting false low-saturation readings when the headband was applied, regardless of body incline. CONCLUSIONS Application of up to 20 mmHg pressure on the forehead pulse oximetry sensor using an elastic tensioning headband significantly reduced reading errors and provided consistent performance when subjects were placed between supine and up to 15 degrees head-down incline (Trendelenburg position).
منابع مشابه
Noninvasive optical method of measuring tissue and arterial saturation: an application to absolute pulse oximetry of the brain.
We present a frequency-domain optical method for real-time noninvasive measurement of absolute tissue and arterial saturation. This method is based on quantitative measurement of the tissue absorption spectrum (for tissue saturation) and of the amplitude of the arterial-pulsation-induced absorption oscillations (for arterial saturation) at eight wavelengths in the range 633-841 nm. We report re...
متن کاملClinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room.
The frequency and nature of spurious pulse oximetry readings were compared using both a conventional pulse oximeter (CPO) and a prototype Masimo signal extraction technology pulse oximeter (Masimo SET). At a university hospital, 50 ASA physical status I-IV adult patients who underwent general or spinal-epidural anesthesia were selected from a group of 250 patients on the basis of high-alarm gen...
متن کاملThe effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.
INTRODUCTION Motion artifact and low perfusion often lead to faulty or absent pulse oximetry readings in clinical practice. OBJECTIVE Determine the impact of motion artifact and low perfusion on newly introduced pulse oximetry technologies during hypoxemic episodes in healthy volunteers. METHODS Five different pulse oximeters from 4 manufacturers (the Datex Ohmeda 3900P; the Agilent; the Ne...
متن کاملEvaluation of a new pulse oximeter sensor.
BACKGROUND A new forehead noninvasive oxygen saturation sensor may improve signal quality in patients with low cardiac index. OBJECTIVES To examine agreement between oxygen saturation values obtained by using digit-based and forehead pulse oximeters with arterial oxygen saturation in patients with low cardiac index. METHODS A method-comparison study was used to examine the agreement between...
متن کاملValidity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia.
During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10), a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt digit sensor (Ivy) to arte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Anesthesiology
دوره 105 6 شماره
صفحات -
تاریخ انتشار 2006