Titration of power output during radiofrequency catheter ablation of atrioventricular nodal reentrant tachycardia.

نویسندگان

  • J J Langberg
  • M Harvey
  • H Calkins
  • R el-Atassi
  • S J Kalbfleisch
  • F Morady
چکیده

Radiofrequency lesions in the anterior, superior aspect of the tricuspid annulus result in selective elimination of fast pathway function in patients with typical atrioventricular (AV) nodal reentry tachycardia. This technique is simple and effective, but has been associated with a significant risk of inadvertent complete AV block. The purpose of this study was to compare the safety and effectiveness of two different techniques for radiofrequency catheter ablation of the fast AV nodal pathway. Initially, a fixed power output was used at each target site. This method was compared retrospectively to a newer technique where power output was gradually incremented at each site. Radiofrequency power was initially applied at 10 watts for 10-15 seconds. If no junctional ectopy or a change in PR interval was seen, power output was incremented by 2 to 4 watts every 10 to 15 seconds up to a maximum of 30 watts. Thirty seven of 38 (96%) patients treated using this incremental power output were cured of their AV nodal reentry tachycardia. None of these patients developed inadvertent complete AV block. In contrast, 92% of historic controls treated with a fixed power output between 20 and 30 watts achieved a primary success and nine of these 89 (10%) historic controls developed inadvertent complete AV block (P = 0.04). There was no difference in the amplitudes of atrial, His, or ventricular electrograms at the effective sites between the two groups.(ABSTRACT TRUNCATED AT 250 WORDS)

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Temperature-controlled slow pathway ablation for treatment of atrioventricular nodal reentrant tachycardia using a combined anatomical and electrogram guided strategy.

AIMS Anatomical and electrogram-guided techniques have been used separately for slow pathway ablation in atrioventricular nodal reentrant tachycardia. The aims of the present study were to analyse electrogram characteristics of target sites and biophysical parameters using a combined anatomical and electrogram-guided technique for temperature-controlled radiofrequency catheter ablation of the s...

متن کامل

Comparison of late results of surgical or radiofrequency catheter modification of the atrioventricular node for atrioventricular nodal reentrant tachycardia.

AIMS Although arrhythmia surgery and radiofrequency catheter ablation to cure atrioventricular nodal reentrant tachycardia differ in technical concept, the late results of both methods, in terms of elimination of the arrhythmogenic substrate and procedure-related new and different arrhythmias, have never been compared. This constituted the purpose of this prospective follow-up study. METHODS ...

متن کامل

Successful radiofrequency catheter ablation of atrioventricular nodal reentrant tachycardia in a patient with dextrocardia due to unilateral pulmonary agenesis: a case report

Radiofrequency catheter ablation of the slow pathway is considered to be the treatment of choice for patients with atrioventricular nodal reentrant tachycardia. We report a 34-year-old female with mirror image dextrocardia due to unilateral pulmonary agenesis who underwent successful slow pathway ablation for typical atrioventricular nodal reentrant tachycardia. Using contrast injection, cardia...

متن کامل

Fast pathway ablation in patients with common atrioventricular nodal reentrant tachycardia and prolonged PR interval during sinus rhythm.

AIMS This study aimed to clarify the safety and efficacy of selective fast pathway ablation in patients with atrioventricular nodal reentrant tachycardia and a prolonged PR interval during sinus rhythm. Such patients have been reported to have an increased incidence of complete atrioventricular block. METHODS AND RESULTS In this study, the earliest retrograde atrial activation during atrioven...

متن کامل

Closed Loop Control

Background The purpose of this study was to evaluate electrode temperatures obtained using a radiofrequency ablation system that incorporates closed loop feedback control to achieve preset target electrode temperatures and to determine if closed loop temperature control results in a lower incidence of developing a coagulum. Methods and Results Two hundred seventy patients underwent catheter abl...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Pacing and clinical electrophysiology : PACE

دوره 16 3 Pt 1  شماره 

صفحات  -

تاریخ انتشار 1993