Biomechanical Properties of a New Anatomical Locking Metal Block Plate for Opening Wedge High Tibial Osteotomy: Uniplane Osteotomy

نویسندگان

  • Seung-Beom Han
  • Ji-Hoon Bae
  • Sung-Jae Lee
  • Tae-Gon Jung
  • Kang-Hee Kim
  • Jae Ho Kwon
  • Kyung Wook Nha
چکیده

PURPOSE The purpose of this study was to evaluate the biomechanical properties of a new anatomical locking metal block plate by comparing the initial biomechanical stability of three different fixation constructs for open wedge high tibial osteotomy (HTO). MATERIALS AND METHODS Sawbones composite tibiae were used to make a 10-mm opening osteotomy model with uniplane technique. The osteotomy was secured with three different types of plates: Group I, new osteotomy plate without a metal block (n=5); Group II, new osteotomy plate with a 10-mm metal block (n=5); and Group III, two short metal block plates (n=5). Single load to failure test and staircase load-controlled cyclical failure test were performed. In the single load to failure test, the yield load, maximum failure load, and the displacement of the osteotomy gap were measured. In the staircase cyclical load to failure test, the total number of cycles to failure was recorded. Failure modes were observed during both single and cyclic load tests. RESULTS Group II showed the highest yield and ultimate loads (1829±319 N, 3493±1250 N) compared to Group I (1512±157 N, 2422±769 N) and Group III (1369±378 N, 2157±210 N, p<0.05). The displacement of the opening gap in Group II (0.34±0.35 mm) was significantly lesser than the other groups (p<0.05). In the staircase cyclical load to failure test, the total number of cycles to failure was 12,860 at 950 N in Group III, 20,280 at 1,140 N in Group I, and 42,816 at 1,330 N in Group II (p<0.05). All the specimens showed complete fracture of the intact lateral sawbones area and slight displacement of the distal fragment of the specimens in the single load to test. None of the specimens showed deformed or broken screws and plates during the single load to test. During the fatigue test with staircase cyclic loading, no fracture of the lateral sawbones area was observed. CONCLUSIONS This study demonstrated that the new anatomical locking metal block plate could provide sufficient primary stability for open wedge HTO. The addition of a metal block to this new plate can increase the stability of the osteotomy compared to the one without a metal block.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Biomechanical Analysis of a Novel Wedge Locking Plate in a Porcine Tibial Model

BACKGROUND The purpose of this study was to analyze biomechanical properties of a novel wedge locking plate in medial open wedge high tibial osteotomy (OWHTO) in a porcine tibial model. METHODS A uniform 8-mm OWHTO was performed in 12 porcine tibiae. Six of them were subsequently fixed with the plate without a wedge, whereas the other 6 were additionally reinforced with a metal wedge of 8 mm....

متن کامل

Supracondylar Osteotomy in Valgus Knee: Angle Blade Plate Versus Locking Compression Plate

  Background: There are few studies comparing the biomechanical properties of angled blade plate and locking compression plates in supracondylar osteotomy. In the current randomized study, we prospectively compared the clinical and radiological outcomes of supracondylar osteotomy using these two plates. Methods: Forty patients with valgus knee malalignment were randomly assigned to two equal nu...

متن کامل

Clinical Outcome of Medial Opening Wedge Osteotomy with T-Locking Plate : Two Years Follow-Up

OBJECTIVE This study was undertaken to determine clinical outcome after medial opening wedge osteotomy with Tlocking plate, with two- year follow up. Twenty-two patients (22 knees) who underwent medial opening wedge osteotomy with T-locking plate (stainless steel 316L, 6 holes) for treatment of varus malalignment of the leg between March 2005 and April 2008 were included in the study. The amoun...

متن کامل

Medial Closing-Wedge Distal Femoral Osteotomy: Fixation With Proximal Tibial Locking Plate.

Distal femoral varus osteotomy is a well-established procedure for the treatment of lateral compartment cartilage lesions and degenerative disease, correcting limb alignment and decreasing the progression of the pathology. Surgical techniques can be performed with a lateral opening-wedge or medial closing-wedge correction of the deformity. Fixation methods for lateral opening-wedge osteotomies ...

متن کامل

High tibial osteotomy: closed wedge versus combined wedge osteotomy

BACKGROUND High tibial osteotomy is a common procedure to treat symptomatic osteoarthritis of the medial compartment of the knee with varus alignment. This is achieved by overcorrecting the varus alignment to 2-6° of valgus. Various high tibial osteotomy techniques are currently used to this end. Common procedures are medial opening wedge and lateral closing wedge tibial osteotomies. The latera...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 26  شماره 

صفحات  -

تاریخ انتشار 2014