The AO dynamic hip screw and the Pugh sliding nail in femoral head fixation.

نویسندگان

  • R H Richards
  • G Evans
  • J Egan
  • J R Shearer
چکیده

We compared, under laboratory conditions, the resistance to cutting out of the AO dynamic hip screw and the Pugh sliding nail. The mean load at cut out, adjusted for bone strength, was 70% greater for the Pugh sliding nail. The reasons for this difference are discussed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Helical blade versus sliding hip screw for treatment of unstable intertrochanteric hip fractures: a biomechanical evaluation.

OBJECTIVE To compare the fixation stability in the femoral head with sliding hip screw versus helical blade designs for unstable, intertrochanteric hip fractures. METHODS A simulated, unstable intertrochanteric hip fracture was created in six pairs of cadaveric femurs. One of each pair was treated using an intramedullary nail with a sliding hip screw (ITST) for femoral head fixation and the o...

متن کامل

Proximal femoral nail for treatment of trochanteric femoral fractures.

PURPOSE To report outcomes of 87 consecutive patients treated with a proximal femoral nail (PFN) for trochanteric femoral fractures. METHODS 17 men and 70 women aged 58 to 95 (mean, 85) years with trochanteric femoral fractures underwent PFN fixation using an intramedullary nail, a lag screw, and a hip pin. Fractures were classified according to the AO system; the most common fracture type wa...

متن کامل

The Warwick Hip Trauma Evaluation One – an abridged protocol for the WHiTE One Study

Fractures of the proximal femur are one of the greatest challenges facing the medical community, constituting a heavy socioeconomic burden worldwide. Controversy exists regarding the optimal treatment for patients with unstable trochanteric proximal femoral fractures. The recognised treatment alternatives are extramedullary fixation usually with a sliding hip screw and intramedullary fixation w...

متن کامل

Dual lag screw cephalomedullary nail versus the classic sliding hip screw for the stabilization of intertrochanteric fractures. A prospective randomized study

This study is a randomized prospective study comparing two fracture fixation implants, the extramedullary sliding hip screw (SHS) and the dual lag screw cephalomedullary nail, in the treatment of intertrochanteric femoral fractures in the elderly. One hundred and sixty-five patients with low-energy intertrochanteric fractures, classified as AO/OTA 31A, were prospectively included during a 2-yea...

متن کامل

Clinical and Functional Outcomes of Treatment for Type A1 Intertrochanteric Femoral Fracture in Elderly Patients: Comparison of Dynamic Hip Screw and Proximal Femoral Nail Antirotation

PURPOSE We aimed to evaluate and compare the clinical and functional outcomes of dynamic hip screw (DHS) and proximal femoral nail antirotation (PFNA) treatment of AO type 1 intertrochanteric fractures in elderly patients. MATERIALS AND METHODS We retrospectively reviewed 194 consecutive patients with type A1 intertrochanteric femoral fractures who were treated with DHS (n=113) or PFNA (n=81)...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 72 5  شماره 

صفحات  -

تاریخ انتشار 1990