Performance verification of the new fully automated Aquios flow cytometer PanLeucogate (PLG) platform for CD4-T-lymphocyte enumeration in South Africa
نویسندگان
چکیده
BACKGROUND The National Health Laboratory Service (NHLS) offers wide-scale CD4 testing through a network of laboratories in South Africa. A new "load and go" cytometer (Aquios CL, Beckman Coulter), developed with a PLG protocol, was validated against the predicate PLG method on the Beckman Coulter FC500 MPL/CellMek platform. METHODS Remnant routine EDTA blood CD4 reference results were compared to results from two Aquios/PLG instruments (n = 205) and a further n = 1885 samples tested to assess daily testing capacity. Reproducibility was assessed using ImmunotrolTM and patient samples with low, medium, high CD4 counts. Data was analyzed using GraphPad software for general statistics and Bland-Altman (BA) analyses. The percentage similarity (%Sim) was used to measure the level of agreement (accuracy) of the new platform versus the predicate and variance (%SimCV) reported to indicate precision of difference to predicate. RESULTS 205 samples were tested with a CD4 count range of 2-1228 cells/μl (median 365cells/μl). BA analysis revealed an overall -40.5±44.0cells/μl bias (LOA of 126.8 to 45.8cells/μl) and %Sim showing good agreement and tight precision to predicate results (94.83±5.39% with %SimCV = 5.69%). Workflow analysis (n = 1885) showed similar outcomes 94.9±8.9% (CV of 9.4%) and 120 samples/day capacity. Excellent intra-instrument reproducibility was noted (%Sim 98.7±2.8% and %SimCV of 2.8%). 5-day reproducibility using internal quality control material (Immunotrol™) showed tight precision (reported %CV of 4.69 and 7.62 for Normal and Low material respectively) and instrument stability. CONCLUSION The Aquios/PLG CD4 testing platform showed clinically acceptable result reporting to existing predicate results, with good system stability and reproducibility with a slight negative but precise bias. This system can replace the faded XL cytometers in low- to medium volume CD4 testing laboratories, using the standardized testing protocol, with better staff utilization especially where technical skills are lacking. Central monitoring of on-board quality assessment data facilitates proactive maintenance and networked instrument performance monitoring.
منابع مشابه
Evaluation of a new single-parameter volumetric flow cytometer (CyFlow(green)) for enumeration of absolute CD4+ T lymphocytes in human immunodeficiency virus type 1-infected Thai patients.
Use of the standard dual-platform flow cytometric method for determination of CD4(+) T-lymphocyte counts, which needs both a flow cytometer (FCM) and hematological analyzer, would inevitably lead to increased variability. The development of new single-platform (SP) FCMs that provide direct CD4(+) T-lymphocyte counts for improved assay precision and accuracy have recently attracted attention. Th...
متن کاملReference ranges for lymphocyte subsets in adults from western India: influence of sex, age and method of enumeration.
BACKGROUND The enumeration of absolute CD4 counts is of primary importance, since therapeutic protocols for HIV1 patients are based on these. AIMS To establish reference ranges for the CD4 and CD8 T-lymphocytes in the Indian population. SETTINGS AND DESIGN Enumeration of absolute numbers and percentages of lymphocyte subsets was performed in 252 healthy adult Indians. METHODS AND MATERIAL...
متن کاملAffordable pediatric CD4 counting by flow cytometry in Malawi
BACKGROUND Rapid expansion of antiretroviral therapy in Malawi has occurred in the relative absence of suitable pediatric CD4 counting facilities. We have recently validated in adults a simplified affordable flow cytometric CD4 counting method, the Blantyre count. There is a need for this technology to transfer to government laboratories run by local staff, and to be validated in children, wher...
متن کاملField evaluation in Chad of community usage of CD4 T lymphocyte counting by alternative single-platform flow cytometry
BACKGROUND Field and community evaluation of the routine usage of CD4 T counting platforms is essential in resource-poor countries for efficient and cost-effective monitoring of HIV-infected adults and children attending health care centers. METHODS We herein addressed the principal issues raised by the implementation of the single-platform, volumetric Auto40 flow cytometer (Apogee Flow Syste...
متن کاملA single platform image cytometer for resource-poor settings to monitor disease progression in HIV infection.
BACKGROUND For resource-poor countries, affordable methods are required for enumeration of CD4(+) T lymphocytes of HIV-positive patients. For infants, additional determination of CD4/CD8 ratio is needed. METHODS We determine the CD4(+) and CD8(+) T lymphocytes as the CD3(+)CD4(+) and CD3(+)CD8(+) population of blood cells. Target cells are CD3-immunomagnetically separated from the whole blood...
متن کامل