Aneurysm rupture after endovascular repair using the AneuRx stent graft.

نویسندگان

  • C K Zarins
  • R A White
  • T J Fogarty
چکیده

OBJECTIVE The purpose of this study was to determine the cause and frequency of aneurysm rupture after endovascular aneurysm repair. METHODS We reviewed each patient who sustained aneurysm rupture among all patients enrolled for endovascular aortic aneurysm repair in phases I, II, and III of the US AneuRx Multicenter Clinical Trial from June 1996 through October 1999. RESULTS A total of 1067 patients were enrolled for endovascular aneurysm repair. The AneuRx stent graft was successfully implanted in 1046 patients (98%). Endovascular repair was unsuccessful in 21 patients (2%); 13 patients (1%) were converted to open aneurysm repair. Among these, two patients (0.2%) sustained aneurysm rupture due to procedure-related instrumentation and underwent open surgical conversion. Aneurysm rupture has occurred in seven patients (0.7%) 3 weeks to 24 months (mean, 16 months) after successful endovascular repair. Four patients survived open surgical repair, and three patients died within 30 days. Overall rupture-related mortality was 0.5% and included late deaths after rupture. Before rupture, two patients had endoleak and aneurysm enlargement, and five patients had no endoleak and no aneurysm enlargement. After aneurysm rupture all seven patients had evidence suggesting that there was poor fixation of the stent graft at the proximal distal, or iliac junction fixation sites. The two patients with endoleak declined recommended open surgical or endovascular repair, which could have prevented aneurysm rupture. In retrospect, the five patients without endoleak could potentially have avoided rupture with better patient selection, better stent graft positioning, or reinforcement of fixation points with stent graft extenders. The probability of no aneurysm rupture for all patients undergoing endovascular repair is 0.996 +/- 0.002 at 1 year and 0.974 +/- 0.011 at 2 years by life table analysis with the longest follow-up of 41 months. CONCLUSION The early risk of aneurysm rupture after endovascular aneurysm repair is low. However, the possibility of rupture persists even in patients with no endoleak after the procedure. Therefore, all patients treated with endovascular aneurysm repair should continue to be monitored after the procedure. Patients with evidence suggesting insecure stent graft fixation should undergo further endovascular treatment or open surgical repair.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Preliminary results of endovascular abdominal aortic aneurysm exclusion with the AneuRx stent-graft.

OBJECTIVES This study evaluated the clinical effectiveness of the Medtronic AneuRx stent-graft in patients with infrarenal abdominal aortic aneurysms (AAAs) who were treated in an endovascular suite. BACKGROUND The use of endovascular stent-graft prosthesis for the treatment of AAAs is receiving increasing attention as an alternative to standard surgical repair. Endovascular treatment of AAAs...

متن کامل

Increased flexibility of AneuRx stent-graft reduces need for secondary intervention following endovascular aneurysm repair.

PURPOSE To evaluate the impact of a change in the manufacturing of the AneuRx stent-graft on the long-term results of endovascular abdominal aortic aneurysm (AAA) repair. METHODS The first 70 AAA patients treated with the AneuRx stent-graft between October 1996 and December 1998 were reviewed. The early stiff bifurcated design (STIFF) was used in 23 patients (mean age 71.7 +/- 9.3 years, rang...

متن کامل

Endoleak as a predictor of outcome after endovascular aneurysm repair: AneuRx multicenter clinical trial.

OBJECTIVE The purpose of this study was to determine whether evidence of blood flow in the aneurysm sac (endoleak) is a meaningful predictor of clinical outcome after successful endovascular aneurysm repair. METHODS We reviewed all patients in Phase II of the AneuRx Multicenter Clinical Trial with successful stent graft implantation and predischarge contrast computed tomographic (CT) imaging....

متن کامل

The AneuRx stent graft: four-year results and worldwide experience 2000.

OBJECTIVE The objective was to review the current results of endovascular abdominal aortic aneurysm repair with the AneuRx stent graft and to determine the effectiveness of the device in achieving the primary objective of preventing aneurysm rupture. METHODS The outcome of all patients treated during the past 4 years in the U.S. AneuRx clinical trial was determined, and the worldwide clinical...

متن کامل

Risk-adjusted outcome analysis of endovascular abdominal aortic aneurysm repair in a large population: how do stent-grafts compare?

PURPOSE To compare differences in the applicability and incidence of postoperative adverse events among stent-grafts used for repair of infrarenal aortic aneurysms. METHODS An analysis of 6787 patients from the EUROSTAR Registry database was conducted to compare aneurysm morphological features, patient characteristics, and postoperative events for the AneuRx, EVT/Ancure, Excluder, Stentor, Ta...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of vascular surgery

دوره 31 5  شماره 

صفحات  -

تاریخ انتشار 2000