Simplified “on‐couch” daily quality assurance procedure for CT simulators

نویسندگان

  • Ruijie Rachel Liu
  • Karl Prado
  • Michael Gillin
چکیده

For most computed tomography (CT) simulators, radiation therapists must remove the flat couch top to perform daily CT quality assurance (QA), and then use separate tools to perform localization-laser QA. This process wastes time and effort and creates the opportunity for accidents to occur. In this study, we tested a simple yet comprehensive daily QA program and phantom that we designed for CT simulators used in radiation oncology that would enable us to use only one tool to perform laser and imaging QA on a flat couch. To construct a modified QA phantom, we attached three adjustable legs and fastened two metric scales (one vertically and one horizontally) to a commercial CT QA phantom. The adjustable legs helped to position the phantom conveniently in the needed position. The two metric scales were used for localization-laser QA, while the phantom body was used for CT imaging QA. We used five different scanners with their designated couches from two manufacturers to evaluate this phantom. Since the couch was scanned with the phantom, we evaluated the couch's effect on image quality. We found that the presence of the couch top changed the uniformity of water's CT number but did not change the visual image resolution; it also produced different, yet reproducible, effects on image quality. The effects were greatest in the section of the phantom closest to the couch top. For a commercial carbon fiber couch top, the variation was within 3 Hounsfield Unit (HU). The effect was couch- and scanner-specific and could be incorporated into the QA acceptability criteria for each CT scanner. By using the proposed program and phantom, we have been able to implement a more thorough QA program while decreasing the amount of effort and time the simulation therapists spend performing laser and imaging QA.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Treatment couch positioning uncertainties using an EPID-based method

Introduction: The accuracy of patient positioning plays an important role for radiotherapy tasks. Short and automated Quality Assurance (QA) programs are becoming one of the challenging tasks in Radiotherapy. The current study, investigates the accuracy of treatment couch positioning with four degrees of freedom (4DoF) using a fast and accurate method based on an image acquired...

متن کامل

An approach for measuring the spatial orientations of a computed‐tomography simulation system

The quality assurance tests for measuring the spatial orientations between tabletop, external patient positioning lasers, couch longitudinal moving direction, and imaging plane in a CT simulation system are a complicated and time-consuming process. We proposed a simple and efficient approach to acquire the angular deviations of spatial orientations between these components. An in-house cross-ji...

متن کامل

Quality assurance for computed-tomography simulators and the computed- tomography-simulation process: Report of the AAPM Radiation Therapy

This document presents recommendations of the American Association of Physicists in Medicine ~AAPM! for quality assurance of computed-tomography~CT! simulators and CT-simulation process. This report was prepared by Task Group No. 66 of the AAPM Radiation Therapy Committee. It was approved by the Radiation Therapy Committee and by the AAPM Science Council. © 2003 American Association of Physicis...

متن کامل

Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: report of the AAPM Radiation Therapy Committee Task Group No. 66.

This document presents recommendations of the American Association of Physicists in Medicine (AAPM) for quality assurance of computed-tomography- (CT) simulators and CT-simulation process. This report was prepared by Task Group No. 66 of the AAPM Radiation Therapy Committee. It was approved by the Radiation Therapy Committee and by the AAPM Science Council.

متن کامل

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi•Art machine

To investigate and evaluate the use of an in-house developed diagnostic software tool using the imaging detector data for a quick daily quality assurance check of the output (dose) and lateral profile (cone) of a tomotherapy Hi*Art system. The Hi*Art treatment system is a radiation therapy machine for delivering intensity modulated radiation therapy (IMRT) in a helical fashion with an integrate...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2009