Quality assurance of a gimbaled head swing verification using feature point tracking
نویسندگان
چکیده
To perform dynamic tumor tracking (DTT) for clinical applications safely and accurately, gimbaled head swing verification is important. We propose a quantitative gimbaled head swing verification method for daily quality assurance (QA), which uses feature point tracking and a web camera. The web camera was placed on a couch at the same position for every gimbaled head swing verification, and could move based on a determined input function (sinusoidal patterns; amplitude: ± 20 mm; cycle: 3 s) in the pan and tilt directions at isocenter plane. Two continuous images were then analyzed for each feature point using the pyramidal Lucas-Kanade (LK) method, which is an optical flow estimation algorithm. We used a tapped hole as a feature point of the gimbaled head. The period and amplitude were analyzed to acquire a quantitative gimbaled head swing value for daily QA. The mean ± SD of the period were 3.00 ± 0.03 (range: 3.00-3.07) s and 3.00 ± 0.02 (range: 3.00-3.07) s in the pan and tilt directions, respectively. The mean ± SD of the relative displacement were 19.7 ± 0.08 (range: 19.6-19.8) mm and 18.9 ± 0.2 (range: 18.4-19.5) mm in the pan and tilt directions, respectively. The gimbaled head swing was reliable for DTT. We propose a quantitative gimbaled head swing verification method for daily QA using the feature point tracking method and a web camera. Our method can quantitatively assess the gimbaled head swing for daily QA from baseline values, measured at the time of acceptance and commissioning.
منابع مشابه
Accuracy verification of infrared marker-based dynamic tumor-tracking irradiation using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000).
PURPOSE To verify the accuracy of an infrared (IR) marker-based dynamic tumor-tracking irradiation system (IR tracking) using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000). METHODS The gimbaled 6-MV C-band x-ray head of the Vero4DRT can swing along the pan-and-tilt direction to track a moving target. During beam delivery, the Vero4DRT predicts the future three-dimensional (3D) target p...
متن کاملControl Simulation and Experimental Verification of Maximum Power Point Tracking Based on RT-LAB (TECHNICAL NOTE)
The maximum power point tracking (MPPT) control in the Photovoltaic system is the key control technology, however present controller has the disadvantages of long development cycle, high cost and complex verification, and there are some disadvantages carrying out totally physical simulation or totally digital simulation of different control algorithms. This paper carried out design of hardware ...
متن کاملPositional accuracy of novel x-ray-image-based dynamic tumor-tracking irradiation using a gimbaled MV x-ray head of a Vero4DRT (MHI-TM2000).
PURPOSE To verify the positional accuracy of a novel x-ray-image-based dynamic tumor-tracking (DTT) irradiation technique using the gimbaled MV x-ray head of a Vero4DRT (MHI-TM2000). METHODS Verification of the x-ray-image-based DTT was performed using three components: a three-dimensional moving phantom with a steel ball target, a laser displacement gauge, and an orthogonal kV x-ray imaging ...
متن کاملMissile Gimbaled Seeker Tracking Using Finite-Horizon State Dependent Riccati Equation
The majority of homing guided missiles use gimbaled seekers. The equations describing seeker gimbal system are highly nonlinear. Accurate nonlinear control of the motion of the gimbaled seeker through the attached DC motors is required. In this paper, an online technique for finite-horizon nonlinear tracking problems is presented. The idea of the proposed technique is the change of variables th...
متن کاملdimensional film dosimetry with GAFCHROMIC films for quality assurance and dosimetric verification of 3D conformal radiotherapy in the presence of heterogeneities
Introduction: The presence of heterogeneities such as air-filled cavities in the head and neck treatment fields region, may result in potential dosimetric disagreement because the losses of charged particle equilibrium. Most of treatments planning systems are not able to predict dose distribution of inhomogeneities region accurately. Therefore, dose calculation algorithms need to...
متن کامل