Treatment plan dosimetric impact after catheter stretch during multi-catheter interstitial breast brachytherapy
نویسندگان
چکیده
Purpose Nylon 6/6 interstitial brachytherapy catheters may stretch when exposed to moisture, mechanical tension, and body temperature. The purpose of this study is to evaluate the magnitude of catheter stretch during a course of multi-catheter interstitial breast brachytherapy for nylon 6/6 treatment catheters, and to assess the impact this has on treatment plan dosimetry. Material and methods Catheters were exposed to water at 37°C for six days and the internal catheter length change (ΔL_W) was measured daily. Additionally, the measured internal catheter length change (ΔL_P), performed prior to each treatment fraction, for seven interstitial breast brachytherapy patients (total of 76 catheters) were retrospectively analyzed. The maximum length change seen in ΔL_P and ΔL_W were introduced as a source positional error to retrospective treatment plans, and treatment plan dosimetry was analyzed. Results ΔL_W and ΔL_P were on average +4.0% and +1.6% of the catheter lengths in water or tissue after 48 hours. Weak correlation was seen between the average ΔL_P per catheter and both the catheter length within tissue (ρ = 0.36, p = 0.0007), and the mid-catheter depth in tissue (ρ = 0.42, p < 0.0001). The D90CTV decreased 1.5% (p < 0.05) and 8.2% (p < 0.05) when the ΔL_P and ΔL_W were introduced to the initial plans. Conclusions Nylon 6/6 catheters stretch during a course of multi-catheter interstitial breast brachytherapy treatment. The observed stretch may affect treatment plan dosimetry, if the catheter internal length is only measured immediately after the insertion. Additional catheter length checks are recommended to verify the actual catheter internal length during the treatment.
منابع مشابه
Multi-catheter interstitial accelerated partial breast irradiation – tips and tricks for a good insertion
Adjuvant radiotherapy is recommended post breast conserving surgery. Accelerated Partial Breast Irradiation (APBI) offers a more attractive shorter course of treatment over 5 days compared to standard conventional external beam radiotherapy, which is often protracted. Multi-catheter interstitial APBI offers excellent dosimetric coverage. This article describes two insertion techniques for multi...
متن کاملMulti-catheter interstitial brachytherapy for partial breast irradiation: an audit of implant quality based on dosimetric evaluation comparing intra-operative versus post-operative placement
PURPOSE The use of multicatheter interstitial brachytherapy (MIB) for accelerated partial breast irradiation (APBI) in early breast cancer (EBC) patients outside the trial setting has increased. Hence, there is a need to critically evaluate implant quality. Moreover, there is a scarcity of reports using an open cavity technique. We report the dosimetric indices of open and closed cavity MIB tec...
متن کاملRadiation dose to the left anterior descending coronary artery during interstitial pulsed-dose-rate brachytherapy used as a boost in breast cancer patients undergoing organ-sparing treatment
PURPOSE To assess dose received by the left anterior descending (LAD) coronary artery during interstitial pulsed-dose-rate brachytherapy (PDR-BT) boost for left-sided breast cancer patients undergoing organ-sparing treatment. MATERIAL AND METHODS Thirty consecutive pT1-3N0-1M0 breast cancer patients boosted between 2014 and 2015 with 10 Gy/10 pulses/hour PDR-BT following a computed tomography...
متن کاملImpact of interfraction seroma collection on breast brachytherapy dosimetry – a mathematical model
PURPOSE Balloon brachytherapy is a widely accepted modality for delivery of accelerated partial breast irradiation (APBI). Our hypothesis was that inter-fraction seroma collection around the balloon surface would have an adverse effect on dosimetry of the target. MATERIAL AND METHODS This is a dosimetric re-planning study using two volumetric models (30 cc and 45 cc) in a Contura(®) multi-lum...
متن کاملDosimetric Improvements in Balloon Based Brachytherapy Using the Contura® Multi-Lumen Balloon (MLB) Catheter to Deliver Accelerated Partial Breast Irradiation
PURPOSE Preliminary dosimetric findings in patients managed with the Contura® Multi-Lumen Balloon (MLB) breast brachytherapy catheter to deliver accelerated partial breast irradiation (APBI) on a multi-institutional phase IV registry trial were reviewed. MATERIAL AND METHODS CT-based 3D planning with dose optimization was performed for all patients. For the study, new ideal dosimetric goals w...
متن کامل