Experimentally induced abutment strains in three types of single-molar implant restorations.
نویسندگان
چکیده
STATEMENT OF PROBLEM The choice of single-molar implant design is difficult because of a lack of controlled, quantitative biomechanical analyses. PURPOSE This study determined the effect of 3 single-molar implant designs on implant strains under a variety of homologous loading conditions. MATERIAL AND METHODS On each implant abutment, 4 strain gauges were placed axially at 90 degrees to each other on the buccal, lingual, mesial, and distal surfaces. Effects of implant design, load location, direction, and magnitude were tested on axial and bending (buccolingual and mesiodistal) strains of 3 single-molar implant designs: (1) single, 3.75-mm (regular) diameter implant, (2) single, 5-mm (wide) diameter implant, and (3) two 3.75-mm diameter (double) implants connected through a single-molar crown. Results were analyzed with ANOVA. RESULTS Variations in loading conditions induced 3-dimensionally complex abutment strains on the tested implant designs. Peak absolute strains in mesiodistal direction were 6493 microepsilon for design 1 and 3958 microepsilon on design 2, and 3160 microepsilon in buccolingual direction on design 3. For all loading conditions, the single 3.75-mm diameter implant consistently experienced the largest strains compared with wide-diameter and double implant designs. Changes in centric contact location affected implant abutment strains differently among the 3 designs. Angulated force direction resulted in larger bending strains. CONCLUSION For single-molar implant designs, an increase in implant number and diameter may effectively reduce experimental implant abutment strains.
منابع مشابه
Stress distribution pattern of screw-retained restorations with segmented vs. non-segmented abutments: A finite element analysis
Background. Screw-retained restorations are favored in some clinical situations such as limited inter-occlusal spaces. This study was designed to compare stresses developed in the peri-implant bone in two different types of screw-retained restorations (segmented vs. non-segmented abutment) using a finite element model. Methods. An implant, 4.1 mm in diameter and 10 mm in length, was placed in t...
متن کاملThe effect of implant angulation and splinting on stress distribution in implant body and supporting bone: A finite element analysis
OBJECTIVE The aim of this study was to investigate the influence of implant crown splinting and the use of angulated abutment on stress distribution in implant body and surrounding bone by three-dimensional finite element analysis. MATERIALS AND METHODS For this study, three models with two implants at the site of mandibular right second premolar and first molar were designed (1): Both implan...
متن کاملAbutment screw loosening in single-implant restorations: a systematic review.
PURPOSE The purpose of this study was to systematically review clinical studies on the incidence of abutment screw loosening in single-implant restorations with different implant-abutment connection geometries. MATERIALS AND METHODS The literature search was conducted using several electronic databases. Specific terms were used for the database search, which spanned the years 1990 to 2006. Th...
متن کاملScrew Joint Stability in Conventional and Abutment-Free Implant-Supported Fixed Restorations.
PURPOSE Procera Implant Bridges (PIBs) do not engage supporting implant shoulders and are fixed using comparably long retention screws. The aim of this in vitro clinical study was to determine the detorque values in PIBs and conventionally fabricated fixed dental prostheses (FDPs). MATERIALS AND METHODS Two groups of screw-retained implant-supported three-unit FDPs (n=10) were fabricated by m...
متن کاملEffect of Three Types of Temporary Luting Cements and Abutment Surface Sandblasting on Retentive Strength of Implant-Supported Fixed Prostheses
Background and Aim: Achieving appropriate retention and easy retrieva bility at the same time is challenging in implant-supported fixed par tial dentures (FPDs). Researchers have always been in search of a temporary cement to improve the retentive strength. The aim of this in vitro study was to evaluate the effect of three different temporary ce ments and sandblasting of the abutment surface on...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of prosthetic dentistry
دوره 84 3 شماره
صفحات -
تاریخ انتشار 2000