EVLA Memo 123 The Effect of the 4-Band Dipoles on L, C, and X Band Performance

نویسنده

  • Rick Perley
چکیده

Test observations show that the deployable 4-band dipoles degrade VLA antenna sensitivity by approximately 7% at L-band, 8% at C-band, and 6% at X-band. The loss at higher frequencies is unknown, but is expected to be less. The dipoles also introduce a variability of 1 – 3% in the antenna cross-polarization at L-band only. This variability will degrade L-band polarimetry capabilities, limiting its accuracy to ∼1%.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

EVLA Memo 109 Wideband Performance of the EVLA L-Band System

The performance of the EVLA L-band feed/receiver across its full 1.0 – 2.0 GHz range is presented. The median efficiency across the band is near 45%, rising by about 4% near 1300 MHz and 2000 MHz, and dropping by a similar value near 1600 MHz. The efficiency drops quickly below 1100 MHz. The variation of focus position with frequency follows predicted values, and an optimum position for wide-ba...

متن کامل

EVLA Memo #152 Characterizing the Sensitivity of the EVLA at L-band

We present sensitivity measurements of the EVLA at L-band between 0.95 and 2 GHz. These are obtained using three EVLA antennas that are equipped with the final, EVLA-compliant L-band receivers. The reported SEFD values meet and/or exceed the EVLA sensitivity requirements over about 70% of the band, but are too high by 12–38% over the frequency range 1425–1750 MHz.

متن کامل

EVLA Memo #119 Wide-Band Sensitivity and Frequency Coverage of the EVLA and VLA L-Band Receivers

We determine the sensitivities of the EVLA and VLA antennas over their full tuning ranges through analysis of the correlation coefficients. Interim EVLA antennas can tune over the full range of 940 through 2200 MHz. The prototype L-band OMT shows good performance over its design frequency range. The betterperforming interim EVLA antennas meet the Project Book sensitivity requirements above ∼115...

متن کامل

EVLA Memo #168 Assessing the Impact of Using Three Cryogenic Compressors on the Performance of the EVLA

We present dewar temperature and system temperature measurements for the eight cryogenically cooled receivers of a subset of four EVLA antennas that have undergone a reconfiguration to use three cryogenic compressors instead of four. We compare the changes in the system temperatures in these antennas with those obtained from a subset of eight other EVLA antennas that have not undergone such a r...

متن کامل

EVLA Memo #188 The Impact of the New Thermal Gap Receiver Assembly on the Sensitivity of the EVLA at C-Band (4–8GHz)

Sensitivity measurements of the EVLA at C-band (4–8GHz) are presented to assess the impact of the receivers with the new thermal gap assembly. The on-the-sky test results are consistent with lab measurements, and show sensitivity improvement across the full frequency range of the C-band receiver. The overall improvement in the sensitivity is ∼9.4%. In the frequency range 4–6GHz, the sensitivity...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2008