A rationale method for evaluating unscrewing torque values of prosthetic screws in dental implants

نویسندگان

  • Felipe Miguel SALIBA
  • Mayra CARDOSO
  • Marcelo Ferreira TORRES
  • Alexandre Carvalho TEIXEIRA
  • Eduardo José Veras LOURENÇO
  • Daniel de Moraes TELLES
چکیده

OBJECTIVES Previous studies that evaluated the torque needed for removing dental implant screws have not considered the manner of transfer of the occlusal loads in clinical settings. Instead, the torque used for removal was applied directly to the screw, and most of them omitted the possibility that the hexagon could limit the action of the occlusal load in the loosening of the screws. The present study proposes a method for evaluating the screw removal torque in an anti-rotational device independent way, creating an unscrewing load transfer to the entire assembly, not only to the screw. MATERIAL AND METHODS Twenty hexagonal abutments without the hexagon in their bases were fixed with a screw to 20 dental implants. They were divided into two groups: Group 1 used titanium screws and Group 2 used titanium screws covered with a solid lubricant. A torque of 32 Ncm was applied to the screw and then a custom-made wrench was used for rotating the abutment counterclockwise, to loosen the screw. A digital torque meter recorded the torque required to loosen the abutment. RESULTS There was a significant difference between the means of Group 1 (38.62 ± 6.43 Ncm) and Group 2 (48.47 ± 5.04 Ncm), with p=0.001. CONCLUSION This methodology was effective in comparing unscrewing torque values of the implant-abutment junction even with a limited sample size. It confirmed a previously shown significant difference between two types of screws.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Screw Joint Stability in Conventional and Abutment-Free Implant-Supported Fixed Restorations.

PURPOSE Procera Implant Bridges (PIBs) do not engage supporting implant shoulders and are fixed using comparably long retention screws. The aim of this in vitro clinical study was to determine the detorque values in PIBs and conventionally fabricated fixed dental prostheses (FDPs). MATERIALS AND METHODS Two groups of screw-retained implant-supported three-unit FDPs (n=10) were fabricated by m...

متن کامل

Influence of different maintenance times of torque application on the removal torque values to loosen the prosthetic abutment screws of external hexagon implants.

PURPOSE To analyze the torque application on prosthetic abutment screws using different maintenance times, to determine an influence on the removal torque values. MATERIALS AND METHODS A total of 40 external hexagon implants, 40 titanium screws, and 40 customized abutments were used. In group 1, the screws received a torque of 30 N·cm by instant torque application; in groups 2, 3, and 4, torq...

متن کامل

Torque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver

Introduction: Some clinicians use a handheld screw driver instead of a torque wrench to definitively tighten abutment screws. The aim of this study was to compare the removal torque of one-piece and two-piece abutments tightened with a handheld driver and a torque control ratchet. Methods: 40 ITI implants were placed in acrylic blocks and divided into 4 groups. In groups one and two, 10 ITI one...

متن کامل

Evaluating the impact of length and thread pitch on the stress distribution in dental implants and surrounding bone using finite element method

 longevity of osseointegrated implants are intensely influenced by biomechanical factors. Control of these factors prevents mechanical complications, which include fracture of screws, components, or materials veneering the framework. In this study, the impact of length and threads pitch of dental implants on the stress distribution and maximum Von Mises stress in implant-abutment complex and ja...

متن کامل

Influence of Prosthetic Screw Material on Joint Stability in Passive and Non-Passive Implant-Supported Dentures

OBJECTIVES This study evaluated the influence of prosthetic screw material on joint stability in implantsupported dentures at two levels of fit. METHODS Ten mandibular implant-supported dentures were fabricated. Twenty cast models were fabricated using these dentures. Four groups (n=10) were tested, according to the vertical fit of the dentures [passive and non-passive] and prosthetic screw m...

متن کامل