Flow-Injection Flame Atomic Absorption Determination of Hexavalent Chromium with On-Line Preconcentration on an Anion Imprinted Polymer
نویسندگان
چکیده
A flow injection preconcentration system for the flame atomic absorption spectrometric determination of hexavalent chromium has been developed. The method employs on-line preconcentration of Cr(VI) on a minicolumn packed with Cr(VI)-imprinted poly(4-vinyl pyridine-co-2-hydroxyethyl methacrylate) placed into a flow injection system. Hexavalent chromium was eluted with a small volume of diluted hydrochloric acid into the nebulizer-burner system of a flame atomic absorption spectrometer. An enrichment factor of 550 and a 3σ detection limit of 0.04 μg⋅L along a sampling frequency of 4 h at a sample flow rate of 3.5 mL⋅min. The relative standard deviation is 2.9% for 1 μg⋅L Cr(VI) (n = 11). The flow injection system proposed has the advantage of being simpler because the use of expensive and sophisticated instruments is avoided. Ease of use, continuous process and selectivity make this method suitable for Cr(VI) determination in different environmental samples such as sea and river waters, soils and sediments.
منابع مشابه
Continuous approach for ultrasound-assisted acid extraction-minicolumn preconcentration of chromium and cobalt from seafood samples prior to flame atomic absorption spectrometry.
A rapid and sensitive method has been proposed for the determination of chromium and cobalt in seafood samples by flame atomic absorption spectrometry combined with a dynamic ultrasound-assisted acid extraction and an on-line minicolumn preconcentration. The use of diluted nitric acid as extractant in a continuous mode at a flow rate of 3.5 mL min(-1) and room temperature was sufficient for qua...
متن کاملApplication of Ion - Imprinted Polymer Synthesized as New Sorbent for Preconcentration and Separation of Thallium (I) and Its Determination by Electrothermal Atomic Absorption Spectroscopy
At this study, thallium (I) ion-imprinted polymer, as a novel sorbent for Tl (I) and comparison with Tl(III) ion-imprinted polymer has been synthesized. This sorbent for preconcentration and solid-phase extraction of Tl(I) ions from aqueous solutions, used. Through preparing the binary complex of Tl(I) ions with 5, 7-dichloroquinoline-8-ol (DCQ) the particles of thallium (I) Ion-Imprinted P...
متن کاملDetermination of chromium(VI) and lead in water samples by on-line sorption preconcentration coupled with flame atomic absorption spectrometry using a PCTFE-beads packed column.
A new time-based flow injection on-line solid phase extraction method for chromium(VI) and lead determination using flame atomic absorption spectrometry was developed. The use of hydrophobic poly-chlorotrifluoroethylene (PCTFE)-beads as absorbent in on-line preconcentration system was evaluated. Effective formation of ammonium pyrrolidine dithiocarbamate complexes and subsequently retention in ...
متن کاملDetermination of chromium(III) and total chromium in water by derivative atomic absorption spectrometry using flow injection on-line preconcentration with a double microcolumn.
A rapid and sensitive method has been proposed for the sequential determination of chromium(III) and total chromium in water samples by flame atomic absorption spectrometry combined with a flow injection on-line preconcentration on a double-microcolumn. The chromium(III) and total chromium in samples were retained on a double-microcolumn with a cation exchange resin, respectively, and eluted di...
متن کاملFlame atomic absorption spectrometric determination of chromium(VI) by on-line preconcentration system using a PTFE packed column.
A new, sensitive and robust time-based flow injection (FI) method for on-line preconcentration and determination of ultra trace amounts of chromium(VI) by flame atomic absorption spectrometry (FAAS) has been elaborated. The sample is initially mixed on-line with ammonium pyrrolidine dithiocarbamate (APDC) and the Cr(VI)-PDC chelate is absorbed quantitatively on a mini-column packed with polytet...
متن کامل