Impact of the vaginal applicator and dummy pellets on the dosimetry parameters of Cs‐137 brachytherapy source
نویسندگان
چکیده
In this study, dose rate distribution around a spherical 137Cs pellet source, from a low-dose-rate (LDR) Selectron remote afterloading system used in gynecological brachytherapy, has been determined using experimental and Monte Carlo simulation techniques. Monte Carlo simulations were performed using MCNP4C code, for a single pellet source in water medium and Plexiglas, and measurements were performed in Plexiglas phantom material using LiF TLD chips. Absolute dose rate distribution and the dosimetric parameters, such as dose rate constant, radial dose functions, and anisotropy functions, were obtained for a single pellet source. In order to investigate the effect of the applicator and surrounding pellets on dosimetric parameters of the source, the simulations were repeated for six different arrangements with a single active source and five non-active pellets inside central metallic tubing of a vaginal cylindrical applicator. In commercial treatment planning systems (TPS), the attenuation effects of the applicator and inactive spacers on total dose are neglected. The results indicate that this effect could lead to overestimation of the calculated F(r,θ), by up to 7% along the longitudinal axis of the applicator, especially beyond the applicator tip. According to the results obtained in this study, in a real situation in treatment of patients using cylindrical vaginal applicator and using several active pellets, there will be a large discrepancy between the result of superposition and Monte Carlo simulations.
منابع مشابه
Simulation of the shielding effects of an applicator on the AAPM TG-43 parameters of CS-137 Selectron LDR brachytherapy sources
Background: The dose rate distribution delivered by a low dose rate 137Cs pellet source, a spherical source used within the source trains of the Selectron gynecological brachytherapy system, was investigated using the MCNP4C Monte Carlo code. Materials and Methods: The calculations were performed in both water and Plexiglas and the absolute dose rate distribution for a single pellet s...
متن کاملInvestigation of Anisotropy Caused by Cylinder Applicator on Dose Distribution around Cs-137 Brachytherapy Source using MCNP4C Code
Introduction: Brachytherapy is a type of radiotherapy in which radioactive sources are used in proximity of tumors normally for treatment of malignancies in the head, prostate and cervix. Materials and Methods: The Cs-137 Selectron source is a low-dose-rate (LDR) brachytherapy source used in a remote afterloading system for treatment of different cancers. This system uses active and inactive sp...
متن کاملDeveloping a Treatment Planning Software Based on TG-43U1 Formalism for Cs-137 LDR Brachytherapy
BACKGROUND The old Treatment Planning Systems (TPSs) used for intracavitary brachytherapy with Cs-137 Selectron source utilize traditional dose calculation methods, considering each source as a point source. Using such methods introduces significant errors in dose estimation. As of 1995, TG-43 is used as the main dose calculation formalism in treatment TPSs. OBJECTIVES The purpose of this stu...
متن کاملSimulation of the shielding effects of an applicator on the AAPM TG-43 parameters of CS-137 Selectron LDR brachytherapy sources
*Corresponding author: Dr. Mohammad Amin Mosleh-Shirazi, Department of Radiotherapy and Oncology, Shiraz University of Medical Sciences, Shiraz, Iran. Fax: +98 711 6474320 E-mail: [email protected] Background: The dose rate distribution delivered by a low dose rate 137Cs pellet source, a spherical source used within the source trains of the Selectron gynecological brachytherapy system, was inv...
متن کاملInvestigation of the Effects of Tissue Inhomogeneities on the Dosimetric Parameters of a Cs-137 Brachytherapy Source using the MCNP4C Code
Introduction: Brachytherapy is the use of small encapsulated radioactive sources in close vicinity of tumors. Various methods are used to obtain the dose distribution around brachytherapy sources. TG-43 is a dosimetry protocol proposed by the AAPM for determining dose distributions around brachytherapy sources. The goal of this study is to update this protocol for presence of bone and air inhom...
متن کامل