Dose Reduction in Automatic Optimization Parameter of Full Field Digital Mammography: Breast Phantom Study
نویسندگان
چکیده
PURPOSE We evaluated the impact of three automatic optimization of parameters (AOP) modes of digital mammography on the dose and image quality. METHODS Computerized Imaging Reference Systems phantoms were used. A total of 12 phantoms with different thickness and glandularity were imaged. We analyzed the average glandular dose (AGD) and entrance surface exposure (ESE) of 12 phantoms imaged by digital mammography in three modes of AOP; namely standard mode (STD), contrast mode (CNT), and dose mode (DOSE). Moreover, exposure factors including kVp, mAs, and target/filter combination were evaluated. To evaluate the quality of the obtained digital image, two radiologists independently counted the objects of the phantoms. RESULTS According to the AOP modes, the score of masses and specks was sorted as CNT>STD=DOSE. There was no difference in the score of fiber among the three modes. The score of image preference was sorted as CNT>STD>DOSE. The AGD, ESE, and mAs were sorted as CNT>STD>DOSE. The kVp was sorted as CNT=STD>DOSE. The score of all test objects in the phantom image was on a downtrend with increasing breast thickness. The score of masses was different among the three groups; 20-21%>30%>50% glandularity. The score of specks was sorted as 20-21%=30%>50% glandularity. The score of fibers was sorted as 30%>20-21%=50% glandularity. The score of image preference was not different among the three glandularity groups. The AGD, ESE, kVp, and mAs were correlated with breast thickness, but not correlated with glandularity. CONCLUSION The DOSE mode offers significant improvement (19.1-50%) in dose over the other two modes over a range of breast thickness and breast glandularity with acceptable image quality. Owning knowledge of the three AOP modes may reduce unnecessary radiation exposure by utilizing the proper mode according to its purpose.
منابع مشابه
Breast dose in conventional and digital mamography
Introduction: The amount of doses received in mammography is one of the biggest concerns in the world. According to some statistics, every 12 minutes a woman dies due to breast cancer. Mammography concerned to detect breast cancer at an early stage of the disease as well as for screening, however due to the risk of exposure, the increasingly and widespread use of mammography fo...
متن کاملAverage Glandular Dose and Entrance Surface Dose in Mammography
In general, dose evaluation for mammography uses average glandular dose (AGD). In mammography, entrance surface dose (ESD) is high, which is well known, but actual clinical data of ESD are not reported so much. This study investigated AGD and ESD for patients whose doses were estimated from 200 mammography examinations using European Organization for Quality-Assured Breast Screening and Diagnos...
متن کاملOptimization of Dose and Image Quality in Full-fiand Computed Radiography Systems for Common Digital Radiographic Examinations
IntroductionA fine balance of image quality and radiation dose can be achieved by optimization to minimize stochastic and deterministic effects. This study aimed in ensuring that images of acceptable quality for common radiographic examinations in digital imaging were produced without causing harmful effects. Materials and MethodsThe study was conducted in three phases. The pre-optimization inv...
متن کاملComparison of contrast detail curves of full field digital with screen film breast phantom images.
In this investigation, we imaged a standard breast phantom and compared the contrast detail curves from a prototype full breast digital mammography system with the corresponding curves for a conventional, analog screen-film system. The full breast digital system exhibited superior contrast detail detectability. The results from this study will be used to plan future clinical evaluations compari...
متن کاملDigital breast tomosynthesis (DBT) versus full field digital mammography (FFDM): comparison of a system performance using a contrast detail phantom
To evaluate performance (image signal to noise ratio) of a digital mammographic system working in 2D planar versus tomosynthesis modality, a contrast detail phantom was developed embedding 1 cm plexyglas, including 49 holes of different diameter and depth, between two layers containing a breast simulating material. The acquisition protocol included 15 low dose projections (reconstructed 1mm-thi...