Evaluation of the Afinion AS100 Point-of-Care Analyzer for Hemoglobin A1c
نویسنده
چکیده
An evaluation of a new Afinion AS100 Analyzer was conducted to assess analytical performance. Precision was estimated by analyzing 2 control and 3 patient samples twice a day for 10 days. Accuracy was established by analysis of 6 samples from the National Glycated Hemoglobin Standardization Program for a 3-day period. Agreement was correlated to a laboratory method, the Variant II Turbo Hemoglobin Testing System, and a point-of-care method, the DCA2000+ Hemoglobin A1c System, using leftover EDTA samples from laboratory analysis (n = 110, range of results = 4.6%Y13.7% HbA1c). The Afinion AS100 Analyzer (0.9%Y1.8% coefficient of variation [CV]) displayed laboratory comparable precision (Variant II Turbo = 1.1%Y1.9% CV) that was superior to the DCA2000+ (2.9%Y3.3% CV) with minimal bias to the National Glycated Hemoglobin Standardization Program target concentrations (G0.2% HbA1c average unit bias or 3.1%). The Afinion AS100 Analyzer had good agreement with both the Variant II Turbo and DCA2000+ with r of greater than 0.9837 and Sy.x of T0.22% and T0.29% HbA1c, respectively. Staff found the analyzer easy to train and use, providing faster results than the DCA2000+ (3 minutes vs 6 minutes). The Afinion AS100 Analyzer will be recommended over the DCA2000+ when requests for future point-of-care HbA1c are made in our health system.
منابع مشابه
Diagnosing diabetes mellitus: performance of hemoglobin A1c point-of-care instruments in general practice offices.
BACKGROUND Hemoglobin A1c (Hb A1c) measurement by hospital laboratory instruments, but not by point-of-care (POC) instruments, has been recommended for use to diagnose diabetes mellitus. We evaluated results from 13 Hb A1c external quality assurance (EQA) surveys over a 6-year period in Norway, from both POC instruments used in general practice (GP) offices and instruments in hospital laborator...
متن کاملAccuracy and precision of the Axis-Shield Afinion hemoglobin A1c measurement device.
BACKGROUND The Afinion HbA1c (Axis-Shield) is a newer point-of-care device for measurement of hemoglobin A1c (A1C) using a boronate affinity method unlike the more commonly used DCA immunoassay method (Siemens Medical Solutions Diagnostics). The Afinion's accuracy and precision, when compared with high-performance liquid chromatography (HPLC) and DCA methods, have not been established in pediat...
متن کاملThree of 7 hemoglobin A1c point-of-care instruments do not meet generally accepted analytical performance criteria.
BACKGROUND In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS Clinical and Laboratory Sta...
متن کاملCorrelation between Standard Clinical Chemistry and Point of Care Testing Analyzers on Hemoglobin A1C Determination
متن کامل
Analysis of the accuracy and precision of the Axis-Shield Afinion hemoglobin A1c measurement device.
Point-of-care (POC) hemoglobin A1c measurement is now used by many physicians to make more timely decisions on therapy changes. A few studies have highlighted the drawbacks of some POC methods, e.g., poor precision and lot-to-lot variability. Evaluating performance in the clinical setting is difficult because there is minimal proficiency testing data on POC methods. In this issue of Journal of ...