Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit
نویسندگان
چکیده
For passive scattering proton therapy systems, neutron contamination is the main concern both from an occupational and patient safety perspective. The Mevion S250 compact proton therapy system is the first of its kind, offering an in-room cyclotron design which prompts more concern for shielding assessment. The purpose of this study was to accomplish an in-depth evaluation of both the shielding design and in-room neutron production at our facility using both Monte Carlo simulation and measurement. We found that the shielding in place at our facility is adequate, with simulated annual neutron ambient dose equivalents at 30 cm outside wall/door perimeter ranging from background to 0.07 mSv and measured dose equivalents ranging from background to 0.06 mSv. The in-room measurements reveal that the H*/D decreases when the distance from isocenter and field size increases. Furthermore, the H*/D generally increases when the angle around isocenter increases. Our results from in-room measurements show consistent trends with our Monte Carlo model of the Mevion system.
منابع مشابه
TOPAS Simulation of the Mevion S250 compact proton therapy unit
As proton therapy becomes increasingly popular, so does the need for Monte Carlo simulation studies involving accurate beam line modeling of proton treatment units. In this study, the 24 beam configurations of the Mevion S250 proton therapy system installed recently at our institution were modeled using the TOolkit for PArticle Simulation (TOPAS) code. Pristine Bragg peak, spread out Bragg peak...
متن کاملCommissioning and initial experience with the first clinical gantry‐mounted proton therapy system
The purpose of this study is to describe the comprehensive commissioning process and initial clinical experience of the Mevion S250 proton therapy system, a gantry-mounted, single-room proton therapy platform clinically implemented in the S. Lee Kling Proton Therapy Center at Barnes-Jewish Hospital in St. Louis, MO, USA. The Mevion S250 system integrates a compact synchrocyclotron with a C-inne...
متن کاملCalculation of Neutron Dose Ratio of Heart, Lung and Liver due to breast cancer Proton Therapy using MCNPX code
Introduction: The proton beam produced in particle accelerators has an appropriate therapeutic potential. In this research, proton therapy of breast cancer is simulated using the MCNPX code in a MIRD phantom, also the contribution of scattered neutron dose during the proton therapy were calculated for the Heart, Lung and Liver. Materials and Methods: For si...
متن کاملAssessment of The Relation Between Energy Of Primary Protons And Undesired Neutron Dose During Proton Therapy By Monte Carlo Method
Introduction: High-energy beams of protons offer significant advantages for the treatment of deepseated local tumors. Their physical depth-dose distribution in tissue is characterized by a small entrance dose and a distinct maximum -Bragg peak- near the end of range with a sharp fall-off at the distal edge. Alongside its advantages there are some point that they need to meticul...
متن کاملEvaluation of Testis functioning during Pelvic Radiotherin the presence of testis lead shield
Introduction: Rectal and bladder cancer are the most common disease among young man. Although External Radiotherapy (EBRT) is one of the most effective treatment options introduced so far, the testis undesired radiation dose increases the risk of infertility and hormonal malfunctioning. The testicles are usually outside of the radiation field, but they can be exposed either to ...
متن کامل