Comparison of the hypothetical 57Co brachytherapy source with the 192Ir source
نویسندگان
چکیده
AIM OF THE STUDY The (57)Co radioisotope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical (57)Co source. MATERIAL AND METHODS A hypothetical (57)Co source was simulated in MCNPX, consisting of an active cylinder with 3.5 mm length and 0.6 mm radius encapsulated in a stainless steel capsule. Three photon energies were utilized (136 keV [10.68%], 122 keV [85.60%], 14 keV [9.16%]) for the (57)Co source. Air kerma strength, dose rate constant, radial dose function, anisotropy function, and isodose curves for the source were calculated and compared to the corresponding data for a (192)Ir source. RESULTS The results are presented as tables and figures. Air kerma strength per 1 mCi activity for the (57)Co source was 0.46 cGyh(-1) cm 2 mCi(-1). The dose rate constant for the (57)Co source was determined to be 1.215 cGyh(-1)U(-1). The radial dose function for the (57)Co source has an increasing trend due to multiple scattering of low energy photons. The anisotropy function for the (57)Co source at various distances from the source is more isotropic than the (192)Ir source. CONCLUSIONS The (57)Co source has advantages over (192)Ir due to its lower energy photons, longer half-life, higher dose rate constant and more isotropic anisotropic function. However, the (192)Ir source has a higher initial air kerma strength and more uniform radial dose function. These properties make (57)Co a suitable source for use in brachytherapy applications.
منابع مشابه
Comparison of the hypothetical 57 Co brachytherapy source with the 192 Ir source Mohammad
tope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical 57Co source. Material and methods: A hypothetical 57Co source was simulated in MCNPX, consisting of an a...
متن کاملCalibration of iridium-192 source by ionization chamber for high dose rate brachytherapy
Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...
متن کاملDetermination of Dosimetric characteristics of a New 192Ir-PDR Brachytherapy Source According to AAPM TG- 43 Protocol using Monte Carlo simulation technique
Introduction: 192Ir is one of the important sources frequently used in brachytherapy. Up to now, a lot of commercial models of this source have been made which Ir-192 has been recently added to them. The aim of the present study is to determine the dosimetric parameters of this new source model based on the recommendations of TG-43(U1) protocol using Monte Carlo simulation tech...
متن کاملComparison the treatment time of cobalt and iridium sources in HDR brachytherapy
Introduction: The use of long-half-life Cobalt source instead of iridium source will be very helpful in HDR brachytherapy for treatment of cervical cancer. The aim of this study was compared the treatment time with cobalt and iridium source in brachytherapy for treatment of cervical cancer using Tandem-ovoid applicator. Materials and Methods: In this study, 19...
متن کاملDetermination of TG-43 Dosimetric Parameters for Photon Emitting Brachytherapy Sources
Objective: Brachytherapy sources are widely used for the treatment of cancer. The report of Task Group No. 43 (TG-43) of American Association of Physicists in Medicine is known as the most common method for the determination of dosimetric parameters for brachytherapy sources. The aim of this study is to obtain TG-43 dosimetric parameters for 60Co, 137Cs, 192Ir and 103Pd brachyt...
متن کامل