Clinical Evaluation of the BD ProbeTecTM Chlamydia trachomatis Q Amplified DNA Assay on the BD ViperTM System With XTRTM Technology
نویسندگان
چکیده
Background: This study evaluated the performance of the BD ProbeTec Chlamydia trachomatis Q (CTQ) Amplified DNA Assay on the BD Viper System with XTR Technology in a multicenter study. Methods: Specimens were collected at 7 geographically diverse clinical sites from 1538 women and men attending sexually transmitted disease, family planning, and obstetrics and gynecology clinics. There were 1465 evaluable participants, 993 women and 472 men. CTQ assay results from female endocervical, self-collected vaginal, male urethral swab specimens, and male and female neat (unpreserved) urine specimens as well as those obtained using the Urine Preservative Transport (UPT) tube for the CTQ assay were compared with patient-infected status (PIS). PIS was determined based on the combined results from Aptima Combo 2 and BD ProbeTec ET CT Amplified DNA Assay. Results: The sensitivity versus PIS for endocervical, vaginal, and both female urine samples was 91.3%, 96.5%, and 93.0%, respectively. The specificity for the same specimen types was 98.3%, 99.2%, and 99.4% (urine neat) and 99.2% (UPT), respectively. The sensitivity versus PIS for male urethral swabs and both male neat and UPT urine were 92.1% and 98%, respectively, with specificities of 98.4%, 99.2%, and 98.1%, respectively. Conclusions: The CTQ assay demonstrated performance characteristics comparable with other commercially available nucleic acid-based tests such as Aptima Combo 2 and BD ProbeTec ET CT-Amplified DNA assay. Vaginal swabs and male urine specimens, the sample types recommended by the Centers for Disease Control for chlamydia screening, both performed at least as well as other sample types evaluated.
منابع مشابه
A two-site analytical evaluation of the BD Viper System with XTR Technology in Nonextracted Mode and Extracted Mode with seeded simulated specimens.
The BD ProbeTec CT Q(x) Amplified DNA Assay and the BD ProbeTec GC Q(x) Amplified DNA Assay (both BD Diagnostics, Sparks, MD) represent two new assays developed for use with the BD Viper System with XTR Technology (BD Diagnostics). These assays were built on the foundation of the former BD ProbeTec ET assays (BD Diagnostics) and its accompanying instrumentation. The study described below compar...
متن کاملComparison of the BD Viper System with XTR Technology to the Gen-Probe APTIMA COMBO 2 Assay using the TIGRIS DTS system for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine specimens.
BACKGROUND Performances of the BD ProbeTec Chlamydia trachomatis (CT)/Neisseria Gonorrhoeae (GC) Q(x) Amplified DNA Assay reagents on a BD Viper System with XTR Technology and APTIMA COMBO 2 Assay reagents on a TIGRIS DTS platform, for detection of both CT and GC were compared. METHODS A total of 1018 first-void urine specimens were tested for the presence of CT and GC DNA using the 2 assays....
متن کاملEvaluation of the BD ProbeTec CTQ Amplified DNA Assay for Detection of the New Chlamydia trachomatis Variant in Sweden
OBJECTIVES The nvCT has a 377 bp plasmid deletion that has caused commonly used nucleic acid amplification tests to generate false negative results. In Swedish counties nvCT has been reported to account for up to 65% of all chlamydia cases in 2007. The aim was to evaluate the CTQ DNA amplification assay for detection of the new Chlamydia trachomatis variant (nvCT) in urine specimens from a labo...
متن کاملMulticenter evaluation of the BDProbeTec ET System for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine specimens, female endocervical swabs, and male urethral swabs.
The performance of the Becton Dickinson BDProbe Tec ET System Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assays (BD Biosciences, Sparks, Md.) was evaluated in a multicenter study. Specimens were collected from 2,109 men and women, with or without symptoms, attending sexually transmitted disease, family planning, and obstetrics and gynecology clinics. Both swab and urine sampl...
متن کاملClinical evaluation of the BD ProbeTec™ Neisseria gonorrhoeae Qx amplified DNA assay on the BD Viper™ system with XTR™ technology.
BACKGROUND The excellent sensitivity and specificity of commercially available nucleic acid amplification tests (NAATs) for the identification of Neisseria gonorrhoeae have been demonstrated. This study evaluated the performance of the BD ProbeTec™ N. gonorrhoeae Q (GCQ) Amplified DNA Assay on the BD Viper™ System with XTR™ Technology in a multicenter study. METHODS Specimens were collected a...