Continuous transcutaneous monitoring of peripheral oxygen and carbon dioxide during infant cardiac surgery

نویسندگان

  • Heping Zhou
  • Hongyan Xiong
  • Chunhu Gu
  • Tao Chen
  • Hailong Zhu
  • Guocheng Sun
چکیده

Close monitoring of the balance between oxygen demand and supply is of great importance during cardiac cardiopulmonary bypass (CPB) surgery. This study was to compare conventional intermittent venous blood gas monitoring with continuous transcutaneous oxygen and carbon dioxide monitoring in infant patients undergoing cardiac surgery with CPB. According to paired data from 29 infant patients undergoing cardiac surgery we found that a positive correlation existed between the two techniques, with a correlation coefficient 0.9021 and 0.8021 for PO2 and PCO2 respectively. It’s concluded that transcutaneous monitoring and intermittent venous blood sampling had good correlation and transcutaneous monitoring may be used conveniently and safely clinically during CPB.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Use of a Combined SpO2/PtcCO2 Sensor in the Delivery Room

Arterial oxygen saturation (SaO(2)) and partial arterial pressure of carbon dioxide (PaCO(2)) are important respiratory parameters in critically ill neonates. A sensor combining a pulse oximeter with the Stow-Severinghaus electrode, required for the measurement of peripheral oxygen saturation (SpO(2)) and transcutaneous partial pressure of carbon dioxide (PtcCO(2)), respectively, has been recen...

متن کامل

New equipment to prevent carbon dioxide rebreathing during eye surgery under retrobulbar anaesthesia.

BACKGROUND Carbon dioxide concentration under ophthalmic drapes increases during eye surgery under local anaesthesia. A new prototype has been designed which combines continuous suction of carbon dioxide enriched air and continuous oxygen insufflation under ophthalmic drapes to prevent carbon dioxide accumulation in spontaneously breathing patients undergoing cataract surgery. METHODS In a pr...

متن کامل

Elective use of high frequency oscillatory ventilation with transcutaneous carbon dioxide monitoring during thoracoscopic diaphragmatic hernia repair

Thoracoscopy, a minimally invasive technique, for congenital diaphragmatic hernia (CDH) repair has been shown to offer significant advantages versus open procedures. However, positive pressure ventilation during thoracoscopy can be challenging. Minimizing lung movement is important to improve surgical visualization, but one-lung ventilation can be difficult in neonates and infants. Ventilation-...

متن کامل

Transcutaneous PCO2 Measurement at Low Temperature for Reliable and Continuous Free Flap Monitoring: Experimental and Clinical Study

BACKGROUND Measurement of transcutaneous oxygen pressure (TcPO2) and transcutaneous carbon dioxide pressure (TcPCO2) has been used for free flap monitoring. Because these values are obtained with sensor probes heated to 44°C, there is potential for low-temperature burns on skin flaps. We measured TcPO2 and TcPCO2 at 37°C in both animals and humans to determine the feasibility and safety of the ...

متن کامل

A Survey of Transcutaneous Blood Gas Monitoring Among European Neonatal Intensive Care Units

Background PCO2 and PO2 are important monitoring parameters in neonatal intensive care units (NICU). Compared to conventional blood gas measurements that cause significant blood loss in preterms, transcutaneous (tc) measurements allow continuous, non-invasive monitoring of blood gas levels. The aim of the study was to survey the usage and opinions among German speaking NICUs concerning tc blood...