Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode

نویسندگان

  • Basem Abdelmalak
  • Narasimhan Jagannathan
  • Faisal D Arain
  • Susan Cymbor
  • Robert McLain
  • John E Tetzlaff
چکیده

Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber paced, dual-chamber sensed, dual response to sensing, and rate modulated) in the coagulating and not cutting mode during a spine procedure. We also discuss the factors affecting intraoperative electromagnetic interference. A 74-year-old man experienced intraoperative electromagnetic interference that resulted in asystole caused by surgical electrocautery in the coagulation mode while the electrodispersive pad was placed at different locations and distances from the operating site (This electromagnetic interference did not occur during the use of the cutting mode). However, because of careful management, the outcome was favorable. Clinicians should be aware that the coagulation mode of electrocautery can cause electromagnetic interference and hemodynamic instability. Heightened vigilance and preparedness can ensure a favorable outcome.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Electrocautery-induced pacemaker malfunction during surgery.

We report a case of electrocautery-induced pacemaker failure that resulted in asystole in a 15-year-old girl scheduled for cardiac surgery. Her pacemaker was converted to the asynchronous mode the night before surgery. Electromagnetic interference from the unipolar electrocautery caused a reduction in the battery voltage, which allowed the digital circuitry, but not the voltage control oscillat...

متن کامل

Erosive Pressure Dermatitis Associated with Cardiac Pacemaker

Erosive dermatitis is a skin reaction associated with implantable cardiac devices (ie, pacemakers and cardioverter defibrillators). We report a 65-year-old man with a history of several cerebrovascular attacks who developed an erythematous crusted erosive patch over the pacemaker site after implantation of two subsequent pacemakers. We suggest that physical irritation may be a trigger factor fo...

متن کامل

Ventricular fibrillation induced by coagulating mode bipolar electrocautery during pacemaker implantation in Myotonic Dystrophy type 1 patient

The occurrence of ventricular fibrillation, induced by bipolar electrocautery during elective dual chamber pacemaker implantation, is reported in a patient affected by Myotonic Distrophy type 1 with normal left ventricular ejection fraction.

متن کامل

Electromagnetic interference between external defibrillator and cardiac resynchronization therapy-pacemaker (CRT-P) devices.

BACKGROUND Implantable heart rhythm devices are susceptible to interference in hospitals where electromagnetic interference (EMI) sources are ubiquitous. CASE DESCRIPTION We report three cases in which EMI from the external defibrillator caused the inability to interrogate Boston Scientific cardiac resynchronization therapy-pacemaker (CRT-P) devices. We have documented interference with the B...

متن کامل

Electromagnetic Interference in a Private Swimming Pool

Although current lead design and filtering capabilities have greatly improved, Electromagnetic Interference (EMI) from environmental sources has been increasingly reported in patients with Cardiac Implantable Electronic Device (CIED) [1]. Few cases of inappropriate intracardiac Cardioverter Defibrillator (ICD) associated with swimming pool has been described [2]. Here we present a case of 64 ye...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 27  شماره 

صفحات  -

تاریخ انتشار 2011