SU-F-T-472: Validation of Absolute Dose Measurements for MR-IGRT With and Without Magnetic Field.

نویسندگان

  • O Green
  • H Li
  • S Goddu
  • S Mutic
  • I Kawrakow
چکیده

PURPOSE To validate absolute dose measurements for a MR-IGRT system without presence of the magnetic field. METHODS The standard method (AAPM's TG-51) of absolute dose measurement with ionization chambers was tested with and without the presence of the magnetic field for a clinical 0.32-T Co-60 MR-IGRT system. Two ionization chambers were used - the Standard Imaging (Madison, WI) A18 (0.123 cc) and the PTW (Freiburg, Germany). A previously reported Monte Carlo simulation suggested a difference on the order of 0.5% for dose measured with and without the presence of the magnetic field, but testing this was not possible until an engineering solution to allow the radiation system to be used without the nominal magnetic field was found. A previously identified effect of orientation in the magnetic field was also tested by placing the chamber either parallel or perpendicular to the field and irradiating from two opposing angles (90 and 270). Finally, the Imaging and Radiation Oncology Core provided OSLD detectors for five irradiations each with and without the field - with two heads at both 0 and 90 degrees, and one head at 90 degrees only as it doesn't reach 0 (IEC convention). RESULTS For the TG-51 comparison, expected dose was obtained by decaying values measured at the time of source installation. The average measured difference was 0.4%±0.12% for A18 and 0.06%±0.15% for Farmer chamber. There was minimal (0.3%) orientation dependence without the magnetic field for the A18 chamber, while previous measurements with the magnetic field had a deviation of 3.2% with chamber perpendicular to magnetic field. Results reported by IROC for the OSLDs with and without the field had a maximum difference of 2%. CONCLUSION Accurate absolute dosimetry was verified by measurement under the same conditions with and without the magnetic field for both ionization chambers and independently-verifiable OSLDs.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Calculation of total dose and dose equivalent distribution in the treatment of lung cancer using MR-guided carbon therapy

Nowadays, in order to improve the accuracy of treatment in radiation therapy, there are many attempts to use magnetic resonance imaging (MRI) due to the advantages of excellent soft tissue contrast and ultra-fast pulse sequences. On the other hand, carbon-ion radiation therapy is developing rapidly due to the benefits of greater relative biological effectiveness (RBE) and the application in the...

متن کامل

Investigation of the dose enhancement effect due to gold nanoparticles at 18 MV radiotherapy using MAGIC-f and Monte Carlo methods Thoraco-Lumbar spinal cord fMRI in 3T Magnetic field

Introduction: Normoxic MAGIC-f polymer gels are established dosimeters used for three dimensional dos uantifications in radiotherapy. The high atomic number nanoparticles such as gold are nov adiosensitizers used to enhance doses delivered to tumors. The aim of this study was t vestigate the effect of gold nanoparticles (GNPs) in enhancing percentage depth doses (PDD ithin the...

متن کامل

Small field size dose-profile measurements using MAGIC- f polymer gel dosimeter and EBT3 film dosimeter

Introduction: In radiotherapy, treatment planning methods are evaluating and transforming every day, so precision in calculation and determination of dose delivery process is performing more important role in radiotherapy than past. In modern radiotherapy treatments, deramatically increase of small fields utilization, causes the polymeric gels has known as the most reliable dos...

متن کامل

Using 3D dosimetry to quantify the Electron Return Effect (ERE) for MR-image-guided radiation therapy (MR-IGRT) applications

Image-guided radiation therapy (IGRT) using computed tomography (CT), conebeam CT, MV on-board imager (OBI), and kV OBI systems have allowed for more accurate patient positioning prior to each treatment fraction. While these imaging modalities provide excellent bony anatomy image quality, MRI surpasses them in soft tissue image contrast for better visualization and tracking of soft tissue tumor...

متن کامل

Effect of uniform magnetic field on dose distribution in the breast radiotherapy

Background: To reduce the dose to normal tissues surrounding the treated breast, a uniform magnetic field was used within a humanoid phantom in breast radiotherapy. Materials and Methods: Monte Carlo simulations were performed with GEANT4, irradiating humanoid phantoms in a magnetic field. To reconstruct phantoms, computed tomography (CT) data slices of four patients were used for the Monte Car...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 43 6  شماره 

صفحات  -

تاریخ انتشار 2016